Page 13 - Shimadzu Xslicer SMX-6010
P. 13

System Configuration  Principles and Features of Planer CT Imaging








 An easy-to-maintain  Capable of Maximum Image Enlargement for Planer Samples

 Shimadzu X-ray generator and various safety mechanisms
 reduce costs and ensure safe use.  Planer CT
                                                                         Scanning large circuit boards
                       X-ray detector                                    with planer CT:
                                                                         The sample is mounted on an XY stage, so it
                                                                         can approach the X-ray generator, enabling
 X-Ray Generator       XY stage                                          maximum image enlargement.

 The Xslicer SMX-6010 features a160-kV open tube and can resolve down to 1 micron.
 Due to the integrated high-voltage transformer, no greasing is necessary.
 The filament replacement process is simplified by automatic adjustments.
 Note: 1  m resolution by JIMA chart analysis.
                                X-ray generator
 Z
 X-Ray Detector  X-ray detector  Tilted  Vertical CT
                                                                         Scanning large circuit boards
 Equipped with a 3 megapixel flat panel detector.  Obstructed             with vertical CT:
                                          Rotary table
                                                         X-ray detector
                                                                         When the sample is brought near the X-ray
 Y                X-ray generator                                        generator to increase the magnification, it
 Five-Axis Stage  Rotated  X                                             will be obstructed by the X-ray generator
                                                                         during CT imaging, making imaging at
 In addition to the X, Y, and Z axes of the sample stage,                maximum magnification impossible.
 the detector tilts and rotates.  XYZ stage  Sample (Large substrate)
 Five-axis movement supports inspections from any angle.  Maximum trajectory for the sample
                      Vertical CT Viewed from above the Rotary Table

 X-Ray Shield Box

 The X-rays are well shielded. (Maximum 1  Sv/h external X-ray leakage)  Observation Examples
 The front door lock mechanism prevents the front door from opening during X-ray emission.
 The interlock mechanism also disables movement of the 5-axis stage when the door is open.
               For samples like circuit boards with planer CT, images with a high magnification can be obtained, unlike with orthogonal CT,
               which is affected by the width of the sample.

 Collision Sensor

 A collision sensor is provided around the X-ray detector to stop the Stage in the event of an emergency (a collision with the
 sample).



 Utilities

 System operation requires only the following: 200 V AC ± 10 %, 1.5 kVA (ground resistance 100   max.)
 There is no need to provide air or cooling water.










                                  Vertical CT                             Xslicer SMX-6010

                                                                                         Xslicer SMX-6010
 12                                                                                             Microfocus X-Ray Inspection System  13