Page 11 - Shimadzu TOC-V Series
P. 11

Extensive Options Expand Functionality  Specifications





 Autosampler  Total Organic Carbon Analyzer TOC-V Series       Autosampler ASI-V


 ASI-V  Even More Functionality and Convenience,  Wet oxidation Total Organic  Vial types  Select from two types:40 mL,125 mL
 Enabling Samples to Be Added During Continuous Measurement  Carbon Analyzer  Number of Vials  40 mL : 68
                                                                                   125 mL : 24
 Features    Model                TOC-VWP         TOC-VWS       Vial septum        with dedicated septum
 Select from two vial types with different capacities to suit your application.  Measurement method  Wet oxidation/NDIR detection  Sample sparging  OK
 ・40 mL vials × 68  ・125 mL vials × 24  Operation method  PC-controlled  Standalone  Dimensions WxDxH(mm)  W370 × D540 × H490 (excluding protrusions)
 All vial types have a well sealed septum that can be easily replaced with the screw   Measured items  TC,IC,TOC,NPOC  Weight  Approx. 14 kg
 cap.
 Changing analysis conditions or adding sample vials during analysis is simple. System   Applicable samples  Aqueous sample  8-Port sampler OCT-1
 operation and carrier gas flow can be automatically stopped on completion of   Combining the ASI-V automatic sample injector  Measurement range  TC:0 to 3,500
             (mg/L)
 measurements.  with a TOC-V Series analyzer    IC:0 to 3,500   Number of OCT-1 Connection  Up to 2 OCT-1 units
 Creating multiple calibration curves and selecting the optimal curves for samples   creates a fully automatic analysis system.  Detection limit  0.5µg/L  Vial types   Any sample container can be used
 with significantly differing concentrations allows these samples to be analyzed in a   Measurement  CV 1.5% max.  8 OCT-1
             accuracy
 single run.  (reproducibility)   (CV 2% max.at 100 mg/L or higher)  Number of  Vials   16 for dual OCT-1
                                                                                    (Sparging is done in the syringe of TOC-V)
 8-Port sampler  Measuring time        TC:Approx. 4 min         Sample spurging    No sparging with OCT-1
                                       IC:Approx. 4 min         Dimensions (mm)    W245 × D245 × H440 (excluding protrusions)
 OCT-1 The Bridge to Ultra-Simplified Automatic Measurement     Weight             Approx. 3.5 kg
             Sample injection         Automatic injection
 Features    Sample injection           350 to 20,400 µL       Solid Sample Combustion
             volume                       variable
 Easy-to-use autosampler does not require special vials.       Unit SSM-5000A
                                       Automatic internal
 Up to 8 samples can be measured with a single OCT-1 unit.  IC pre-treatment  acidification and sparging
 Up to 16 samples can be measured by adding a second OCT-1.  Automatic dilution  Dilution factor 2 to 50  TC:Catalytically aided combustion oxidation at 900°C.
 Commercially available stirrers can be used. (Stirrers are sold separately.)  Carrier gas  High purity nitrogen  Method  IC:Pre-acidification, oven temperature: 200°C.
 Samples can be added during continuous measurement.            Measured Item      TC,IC,TOC
             Gas pressure                 300 kPa
 * Select either ASI-V or OCT-1. They cannot be used simultaneously.
                                     Approx. 3,000 L/month                         TC:0.1 to 30 mg carbon
             Gas consumption                                    Measuring Range*    (1~20 g carbon in high sensitivity measurement)
                              (operating conditions: 8 hours/day × 5 days/week)    IC:0.1 to 20 mg carbon
 Solid Sample Combustion Unit  Operating keys  Use PC  Built-in  Sample Amount*    1gram-aqueous content < 0.5 g
             Display               Use PC        Built-in LCD   Analysis Time*     5 to 6 minutes at a gas rate
 SSM-5000A Capable of TOC Measurements in Solid Samples  Printer  PC printer  Built-in  99.9% O2 at the 500 mL/min.
             Ambient temperature range    5 to 35°C             Carrier Gas        High-purity O2 Gas is required for high
 Features                                                                          sensitivity measurement.
             Power supply         AC100~127 V±10%, MAX350 VA
 Measurement of maximum 1 g samples with up to 30 mg carbon content reduces   AC220~240 V±10%, MAX350 VA
 weighing errors, and errors due to uneven distribution of the sample carbon   Dimensions (mm)  W440 × D560 × H460 (excluding protrusions)  Power Requirement  100~127 V or
                                                                                   220~240 V as ordered.
 content.    Weight                     Approx. 40 kg                              700 VA, 50/60 Hz
 A single TOC analyzer unit can handle a range from several 10 ppb TOC in ultrapure   Dimensions WxDxH(mm)  W450 × D656 × H290
 water to several 10% TOC in solid samples.                     Weight             Approx. 30 kg
 Measurement of inorganic carbon (carbonate) in solid samples.  In addition to aqueous samples, carbon   *Will vary with sample type and measurement condition.
 Measurement of aqueous samples containing large quantities of suspended   measurements can also be performed on soil,
 substances.  sludge, sedimentation, and other solid samples.
 By swabbing, the carbon in attached residues
 Simply change the onscreen settings to switch between aqueous sample   can be measured for cleaning validation.
 measurement with the TOC-V, and solid sample measurement using the SSM-5000A.

 【GMP Cleaning Validation Using the Swab Method】
 Swab sampling  Transfer to sample boat  TC measurement







 Residue evaluation using swab sampling with quartz microfiber filter paper and direct combustion carbon analysis
  (Please contact your local Shimadzu representative for further information.)

                                                                                               TOC-V Series