Extensive Options Expand Functionality Specifications
Autosampler Total Organic Carbon Analyzer TOC-V Series Autosampler ASI-V
ASI-V Even More Functionality and Convenience, Wet oxidation Total Organic Vial types Select from two types:40 mL,125 mL
Enabling Samples to Be Added During Continuous Measurement Carbon Analyzer Number of Vials 40 mL : 68
125 mL : 24
Features Model TOC-VWP TOC-VWS Vial septum with dedicated septum
Select from two vial types with different capacities to suit your application. Measurement method Wet oxidation/NDIR detection Sample sparging OK
・40 mL vials × 68 ・125 mL vials × 24 Operation method PC-controlled Standalone Dimensions WxDxH(mm) W370 × D540 × H490 (excluding protrusions)
All vial types have a well sealed septum that can be easily replaced with the screw Measured items TC,IC,TOC,NPOC Weight Approx. 14 kg
cap.
Changing analysis conditions or adding sample vials during analysis is simple. System Applicable samples Aqueous sample 8-Port sampler OCT-1
operation and carrier gas flow can be automatically stopped on completion of Combining the ASI-V automatic sample injector Measurement range TC:0 to 3,500
(mg/L)
measurements. with a TOC-V Series analyzer IC:0 to 3,500 Number of OCT-1 Connection Up to 2 OCT-1 units
Creating multiple calibration curves and selecting the optimal curves for samples creates a fully automatic analysis system. Detection limit 0.5µg/L Vial types Any sample container can be used
with significantly differing concentrations allows these samples to be analyzed in a Measurement CV 1.5% max. 8 OCT-1
accuracy
single run. (reproducibility) (CV 2% max.at 100 mg/L or higher) Number of Vials 16 for dual OCT-1
(Sparging is done in the syringe of TOC-V)
8-Port sampler Measuring time TC:Approx. 4 min Sample spurging No sparging with OCT-1
IC:Approx. 4 min Dimensions (mm) W245 × D245 × H440 (excluding protrusions)
OCT-1 The Bridge to Ultra-Simplified Automatic Measurement Weight Approx. 3.5 kg
Sample injection Automatic injection
Features Sample injection 350 to 20,400 µL Solid Sample Combustion
volume variable
Easy-to-use autosampler does not require special vials. Unit SSM-5000A
Automatic internal
Up to 8 samples can be measured with a single OCT-1 unit. IC pre-treatment acidification and sparging
Up to 16 samples can be measured by adding a second OCT-1. Automatic dilution Dilution factor 2 to 50 TC:Catalytically aided combustion oxidation at 900°C.
Commercially available stirrers can be used. (Stirrers are sold separately.) Carrier gas High purity nitrogen Method IC:Pre-acidification, oven temperature: 200°C.
Samples can be added during continuous measurement. Measured Item TC,IC,TOC
Gas pressure 300 kPa
* Select either ASI-V or OCT-1. They cannot be used simultaneously.
Approx. 3,000 L/month TC:0.1 to 30 mg carbon
Gas consumption Measuring Range* (1~20 g carbon in high sensitivity measurement)
(operating conditions: 8 hours/day × 5 days/week) IC:0.1 to 20 mg carbon
Solid Sample Combustion Unit Operating keys Use PC Built-in Sample Amount* 1gram-aqueous content < 0.5 g
Display Use PC Built-in LCD Analysis Time* 5 to 6 minutes at a gas rate
SSM-5000A Capable of TOC Measurements in Solid Samples Printer PC printer Built-in 99.9% O2 at the 500 mL/min.
Ambient temperature range 5 to 35°C Carrier Gas High-purity O2 Gas is required for high
Features sensitivity measurement.
Power supply AC100~127 V±10%, MAX350 VA
Measurement of maximum 1 g samples with up to 30 mg carbon content reduces AC220~240 V±10%, MAX350 VA
weighing errors, and errors due to uneven distribution of the sample carbon Dimensions (mm) W440 × D560 × H460 (excluding protrusions) Power Requirement 100~127 V or
220~240 V as ordered.
content. Weight Approx. 40 kg 700 VA, 50/60 Hz
A single TOC analyzer unit can handle a range from several 10 ppb TOC in ultrapure Dimensions WxDxH(mm) W450 × D656 × H290
water to several 10% TOC in solid samples. Weight Approx. 30 kg
Measurement of inorganic carbon (carbonate) in solid samples. In addition to aqueous samples, carbon *Will vary with sample type and measurement condition.
Measurement of aqueous samples containing large quantities of suspended measurements can also be performed on soil,
substances. sludge, sedimentation, and other solid samples.
By swabbing, the carbon in attached residues
Simply change the onscreen settings to switch between aqueous sample can be measured for cleaning validation.
measurement with the TOC-V, and solid sample measurement using the SSM-5000A.
【GMP Cleaning Validation Using the Swab Method】
Swab sampling Transfer to sample boat TC measurement
Residue evaluation using swab sampling with quartz microfiber filter paper and direct combustion carbon analysis
(Please contact your local Shimadzu representative for further information.)