Page 5 - Shimadzu Tm Analysis System
P. 5

Reliable Data Integrity  Functionality for Satisfying Various Needs





 System Configuration  8-Cell Micro Multi-Cell




 Tm analysis system consists of the following.  This micro multi-cell can be used to measure eight
                                                                   1 mm Optical Path Length  10 mm Optical Path Length
 • UV-Vis Spectrophotometer  samples at the same time using the same   NEW
                                                                   Micro Multi-Cell   Micro Multi-Cell
 • TMSPC-8i 8-cell thermoelectrically temperature controlled cell holder  measurement conditions.
 • LabSolutions UV-Vis Tm software  In addition to previous micro multi-cell models
                                                             3.000
               with a 10 mm optical path length (with a 100  L                                    60 µM
               minimum sample volume) for satisfying         2.400
               high-sensitivity measurement needs, a new model   1.800                            40 µM
               with a 1 mm optical path length (with a 10  L   Absorbance at 260 nm (Abs)
               minimum sample volume) for measuring          1.200                                20 µM
               micro-quantities of expensive samples has been   0.600                             12 µM
               added to the lineup. Cells with a 1 mm optical path                                 9 µM
                                                             0.000
 ™
 NEW  TMSPC -8i  UV-Vis  NEW  LabSolutions  UV-Vis Tm  length can result in sample evaporation problems,   40.00  50.00  60.00  70.00  80.00  90.00
 ™
               but a new sealing method*, which suppresses the                 Temperature
               evaporation of samples by sealing the top side of     Melting Curve of Nucleic Acid (M13 Primer)
                                                                 (Using Micro Multi-Cell with 1 mm Optical Path Length)
               the cell, enables reliable measurement of samples
 Data Integrity
               with high melting temperatures.
               *The sealing method was supported by Professor Junji Kawakami
 Tm analysis systems comprise a spectrophotometer, TMSPC-8i 8-cell thermoelectrically temperature-controlled cell   (Department of Nanobiochemistry, Faculty of Frontiers of Innovative
               Research in Science and Technology (FIRST), Konan University, Japan).
 holder, and LabSolutions UV-Vis Tm software. They can automatically determine Tm values based on melting curve
               This was supported by AMED under Grant Number JP21ae0121022,
 measurements using the average method or derivative method. LabSolutions UV-Vis Tm can provide the industry’ s   JP21ae0121023, JP21ae0121024 (Project leader: Satoshi Obika).
 highest data integrity levels when linked to a LabSolutions DB/CS system, which has an extensive track record from use
 with LC, GC, and many other analytical instruments.
               Real-Time Transfer Function and Thermodynamic Parameter Analysis

 Client PC  Laboratory or office  LabSolutions server
               LabSolutions UV-Vis Tm can transfer temperature,
                                                                 Temperature
               absorbance, and time values to an Excel  file in real          Real-time data
                                            ®
                                                                               transfer
               time during measurements.
                                                                 Absorbance
 Acquisition controller PC  By transferring Tm values obtained from samples                X
               with different concentration levels to commercial   Time
               spreadsheet software, changes in Gibbs free energy
               values used as an index for drug activity levels or
               thermodynamic properties, such as entropy or
               enthalpy, can be analyzed easily.

 Note: LabSolutions DB UV-Vis: This is a stand-alone database management system which analysis software on a single PC.
 LabSolutions CS system: A server, client, and acquisition controller are connected on the network. The data of all devices are centrally managed by the
 database built on the server. User and permission management is centrally managed with this system.


 4                                                                                                                 5