Page 15 - Shimadzu SALD-2300
P. 15
System Structure
Powder samples can be measured without preparation.
Cyclone Injection Type Dry Measurement Unit SALD-DS5S
A cyclone type sample suction mechanism has been developed. Measurement Results
Strong double dispersion process of suction and injection can be used. Q3(%) q3(%) Hard flour Dispersion of flour by water is so difficult that organic
Measurement with high precision, high sensitivity, high reproducibility and high resolution 20 Soft flour solutions such as IPA must be used for wet
■Sample to be applied 80 15 measurement.
・ Easily dissolved samples (medicine, powder foods) Normalized Particle Amount (Diff) 100 60 Cyclone injection mechanism can disperse flour in air
・ Easily agglomerated samples (magnetized particles) 40 10 Normalized Particle Amount (Diff) easily, enabling dry measurement with good
■Features 5 reproducibility.
・ Using three types of sample suction mechanism (cyclone type, one shot type, hand shot type) and four types of dispersion nozzle, the optimum combination 20 Moreover, the sample can be gathered into a vacuum
can be selected to perform measurements based on the sample properties and amount. 0 0
0.01 0.05 0.1 0.5 1 5 10 50 100 500 1000 5000 cleaner and disposed of easily.
・ When the cyclone type is used, the sample loaded in the special sample holder is sucked in while being rotated, then injected from the nozzle and measured. Particle Diameter (µm)
In addition, since a sample holder is used, there is no scattering of the sample nor do the operator's hands become soiled.
・ When the one shot type is used, putting the sample into a small hopper is the only operation required for measuring. This type is suitable for a small sample Q3(%) q3(%) Q3(%) q3(%)
amount. (option) 10 10
・ When the hand shot type is used, the sample can be sucked directly from the beaker or chartula for measurement. (option) 80 8 80 8
Sample Suction Normalized Particle Amount (Diff) 100 60 6 Normalized Particle Amount (Diff) Normalized Particle Amount (Diff) 100 60 6 Normalized Particle Amount (Diff)
Mechanism 40 4 40 4
One Shot Type (option) 20 2 20 2
Measurement Unit
Hopper 0 0.01 0.05 0.1 0.5 1 5 10 50 100 500 1000 5000 0 0 0.01 0.05 0.1 0.5 1 5 10 50 100 500 1000 5000 0
Wing Sensor Particle Diameter (µm) Particle Diameter (µm)
Ejector Sensor for Forward Scattered Light 2 types of gastrointestinal medicines
Air Inlet to Prevent Backflow
Scattered Light Dissolubility depends on particle size distribution and may adversely impact the measurement of pharmaceuticals. Dry measurement is
Hand Shot Type (option) 4 Types Condensing Lens necessary to measure pharmaceuticals with dissolution effects.
Suction Tube Ejector Injection Nozzle Sample Particles
Injection Suction
Laser Beam
Hand Switch Air Valve PC
Sample can be injected
Suction Nozzle
only under the operation
of a vacuum cleaner Pressure Sensor to
Confirm the Operation
Cyclone Type
Pressure Regulator of Dust Collector Dust
(PC Control) Collector
Compressor
(Compressed Air
Source)
・ When compressed air, including the sample, flows through the injection nozzle, the shape,
area, and direction of the cross section is changed in order to obtain the large changes in
volume, pressure and direction of air flow. Therefore, agglomerates can be strongly
dispersed into air.
・ In order to obtain the optimum dispersion based on the properties of the sample one of
Standard dispersion nozzle (Type 1) Sample four differently shaped nozzles can be selected.
Section-2 Injection ・ Magnetized particles, which are easy to cohere in liquid, can be strongly dispersed into
Section-1 Section-4
Section-3 air using the type 1 injection nozzle. Therefore, accurate measurement results can be
3 mm dispersion nozzle (Type 2 * ) 1 obtained.
Section-2 ・ Using a type 4 dispersion nozzle the dispersion force is weakened, enabling measurements
Section-1 Section-4
Section-3 to be made without destroying fragile particles.
2.5 mm dispersion nozzle (Type 3 * ) 1 ・ Using type 2 and 3 dispersion nozzles the dispersion force is strengthen, that is suitable
Section-2 for the dispersion of fine particles.
Section-1 Section-4
Section-3 ・ When the cyclone type is used, the double dispersion process of suction and injection
Straight nozzle (Type 4 * ) 2 enables measurement with good reproducibility.
Section-2 Section-4
Section-1 Section-3 * Type 2 and Type 3 are set options * Type 4 is an option SALD-2300
2
1
14 Laser Diffraction Particle Size Analyzer 15