Page 17 - Shimadzu SALD-2300
P. 17
Hardware Specifications Software Specifications
General Specifications WingSALD II
Measurement principle Laser Diffraction Method Measurement and Data Display Functions
Measurement range SALD-MS23: 17nm(0.017µm) to 2500µm Measurement of Particle Size Distribution Allows measurements using measurement assistant function (interactive process based on SOP)
SALD-BC23: 17nm(0.017µm) to 400µm Refractive Index Setting Automatic refractive index calculation function (LDR method: Light Intensity Distribution Reproduction Method)
SALD-HC23: 30nm(0.030µm) to 280µm makes setting the refractive index easy.
SALD-DS5: 300nm(0.3µm) to 2500µm Real-Time Display Particle size distribution/light intensity distribution simultaneous display
Note 1: The measurement range varies according to the shape etc. of the particle.
Diagnostics/Adjustments Self-diagnostic function and cell check function
Measurement Unit: SALD-2300 Recalculation of Particle Size Distribution Batch recalculation of max. 200 distributions
Light source Red Semiconductor laser (Wavelength 680 nm) Display of Particle Size Distribution Data Displays overlay of max. 200 distributions
Light detector Detector elements for UV semiconductor laser Display of Light Intensity Distribution Displays overlay of max. 200 distributions
Total 84 elements (78 forward, 1 side, 5 back) Statistical Data Processing Max. 200 sets of data (also allows overlaying max. 200 data sets)
System Compliance Class 1 Laser Product, CE compliant
Time-Series Processing Max. 200 sets of data
Required power supply 115 or 230 VAC as ordered 100 VA
3-Dimensional Graphing Max. 200 sets of data
Dimensions & weight W680mm×D280mm×H430mm, 31kg
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) Data Transfer via Clipboard [Image Output]: Outputs entire data sheet or graph only.
Note 2: Reference sample and USB cable (2 m) supplied as standard [Text Output]: Outputs summary data, particle size distribution data, or light intensity distribution data.
Data Sorting Sorts by file name, sample ID, sample number, or refractive index
Sampler: SALD-MS23 Output Conditions
Dispersing bath Capacity: 100~280cm 3 Particle Size (µm) Divisions Fixed 51 or 101 divisions User settable 51 divisions
Sonicator Frequency about 32 kHz, output about 40 W
Liquid Pump Radial pump, maximum flow rate 2000cm 3 /min Particle Amount (%) Divisions Fixed 51 divisions User settable 51 divisions
Liquid Pump Material Stainless (SUS 304, SUS 316), Tetrafluoroethylene (PTFE), Distribution Basis Count, length, area, or volume
Perfluoroelastmor (FEP) or Kalrez, Thermoflon Pascal (inside) Expression of Cumulative Distribution Oversized or undersized
Liquid Supply Pump Diaphragm pump, maximum flow rate 750cm 3 /min Expression of Frequency Distribution q, q / Δ×, q / Δlog ×
Liquid Supply Pump Material Tetrafluoroethylene, polyvinyl dene fluoride Smoothing Levels 10 levels
Flow Cell Quartz glass Distribution Function Fitting Rosin-Rammler distribution, logarithmic Gaussian distribution
Required power supply 115 or 230 VAC as ordered, 200 VA Data Shifting ±10 levels
Dimensions & weight W390mm×D520mm×H430mm,18kg Report Function Single data sets (6 templates), overlaid data (5 templates), statistical data, time-series data,
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)
or 3D data can be selected and output using batch processing
Note 3: USB cable (2 m) supplied as standard
Data Analysis Functions
Batch Cell: SALD-BC23 Scattering Angle Evaluation Function Evaluates scattering characteristics within micro angle regions for samples such as optical films and sheets.
Cell Material Quartz glass Data Emulation Functions Emulates measurement results from other instruments and measurement principles, using SALD series
Required Liquid Volume Approx. 12 cm 3 measurement results.
Stirrer Mechanism Up-and-down movement of blade Mixture Data Simulation Function Simulates particle size distributions using any mixture ratio of multiple particle size distributions.
Dimensions & weight W100mm×D120mm×H140mm, 0.8kg Data Connection Function Combines two particle size distributions with different measurement ranges at any particle size point to
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)
create a single particle size distribution.
Continuous Measurement Function Continuously measures changes in particle size distributions and particle diameters over time,
High-concentration Sample Measurement System: SALD-HC23S at intervals as short as one second, and saves the results.
Cell Material Borosilicate glass
Required Liquid Volume Approx. 0.15 cm 3 Note 4: The LDR method automatically calculates an appropriate refractive index based on consistency between the actual measured light intensity distribution
and one reproduced (recalculated) from particle size distribution data. This method was developed by Shimadzu and published in two technical papers.
Dimensions & weight W20mm×D100mm×H9mm, 0.2kg It is sometimes called the "Kinoshita Method", in academic communities, after the name of Shimadzu’s engineer.
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)
Cyclone Injection Type Dry Measurement Unit: SALD-DS5S
Sample suction types Cyclone type / One shot type / Hand shot type PC Requirements The software is included standard on a CD-R with the SALD system (optical system).
Suction nozzle Can be selected from 4 types Install the software on a PC that meets the following specifications.
Sampling Unit Specifications
System Cyclone type OS Windows 10 (32/64bit)
Communication method USB (PC control) CPU Core i5, i7
Required power supply 115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor) MEMORY 4GB min.
Dimensions & weight W 240×D 310×H 210 mm, 10 kg HDD Min. 1 GB of free space required.
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) CD-ROM Drive Required for software installation
Pressure Regulator Specifications USB Port Unit Name Required USB port
Primary pressure 0.6 to 0.8 MPa SALD-2300 1 port
Secondary pressure 0.05 to 0.5 MPa
SALD-BC23 0
Filtration rating Removing particles of 5 µm or larger
Connection to air source Tube of 6 mm outside diameter SALD-MS23 1 port
Communication method USB (PC control) SALD-HC23S 0
Required power supply 115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor) SALD-DS5S 2 ports
Dimensions & weight W 130×D 223×H 233 mm, 3 kg Printer 1 port
Operating Environment Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) Display SXGA (1280×1024 pixels) min.
Printer Must be compatible with operating system.
Requirements for compressor and dust collector *1: Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
Compressor Output : 0.4 kW *2: Other company names and product names mentioned in this document are registered trademarks or trademarks of their respective companies.
Minimum pressure : Approx. 7 kgf/cm 2 (approx. 0.69 MPa) *3: The TM and ® symbols are omitted in this document.
Air discharge : 45 L/min
Tank volume : Approx. 30 L
Dust collector Type : Vacuum cleaner (Paper pack type)
Dust collecting efficiency : More than 99% for 0.3 µm particles
Capacity : 2.0 m 3 /min or more
Vacuum : Approx. 2000 mm Aq or lower
Suction hose diameter : Approx. 32 mm
SALD-2300
16 Laser Diffraction Particle Size Analyzer 17