Page 12 - Shimadzu RF-6000
P. 12

High Sensitivity, High Stability and High Speed







              High-Sensitivity Measurements




            ■ High-sensitivity measurements can be performed with S/N ratio of 1,000:1 (RMS) or 350:1 (P–P).
            ■ Improved lower limit of quantitation. Measures concentrations up to 1 × 10 −13  mol/L (fluorescein).

            A redesigned optical system and signal processing system achieve the highest S/N ratio levels in its class.
            Even extremely dilute samples can be measured accurately.

            The RF-6000 can measure fluorescence spectra from fluorescein concentrations as low as 1 × 10 −13  mol/L. Furthermore, due to an auto-gain
            control function that ensures measurements are performed using optimal measurement parameters, the system can perform accurate and highly
                                                                    −7
            quantitative measurements over a wide six-order dynamic range, from 10 −13  to 10  mol/L.





                                                                                                 10 −10  mol/L
                                                  10 −7  mol/L

                                                                           10 −12  mol/L
                                                                                                 10 −11  mol/L
                                                  10  mol/L
                                                    −8
                                                                                        10 −13  mol/L
                                         −7
                                  10 −13  to 10  mol/L                          10 −13  to 10 −10  mol/L
                                                 Fluorescence Spectra of Fluorescein


                                           Excitation: 485 nm
                                           Florescence: 515 nm
                                            2
                                           R =1.000













                                                        Calibration Curve





              High-Speed Measurements




            Enables ultrafast scanning at speeds up to 60,000 nm/min. All wavelength regions can be measured in only one second. 3D fluorescence spectra
            can also be measured quickly.