Page 10 - Shimadzu RF-6000
P. 10
Environmental
Measuring Oil in Water - ASTM D5412
The American Society for Testing and Materials (ASTM) testing standard D 5412 specifies testing for polycyclic aromatic hydrocarbons contained in
water as oil. In this example, the target substance was separated from a solution prepared with a mixture of five polycyclic aromatic hydrocarbons
using synchronized scan*. Fig. 1 is a fluorescence spectrum of the mixture of polycyclic aromatic hydrocarbones. Fig. 2 is a synchronized scan
spectrum of Benzo[a]pyrene. Fig. 3 is a synchronized scan spectrum of the mixture. The fluorescence peak of Benzo[a]pyrene was not
distinguished in the fluorescence spectrum of the mixture; however, it was clearly separated on the synchronized scan spectrum.
* Synchronized scan
Synchronized scan mode simultaneously scans samples using both an excitation monochromator and a fluorescence monochromator that are
offset by fixed wavelength intervals. Sharp fluorescence peaks can be detected for target substances even if multiple types of components are
mixed together.
Fluorescence Intensity
Fig. 1 Fluorescence Spectrum of the Mixture of Polycyclic Aromatic Hydrocarbons (Ex 300 nm)
Fluorescence Intensity Fluorescence Intensity
Fig. 2 Synchronized Scan Spectrum of Fig. 3 Synchronized Scan Spectrum of
Benzo[a]pyrene (Offset 6 nm) the Mixture (Offset 6 nm)
10