Page 10 - Shimadzu Prominence
P. 10

Autosamplers






            SIL-20A / 20AHT
            Supporting High-Throughput Analysis
            The SIL-20A is a total-volume injection-type Autosampler that enables high-speed
            injection and multi-sample processing. It was designed to ensure greater stability,
            with improved durability attained through modifications in valves and sample loops.

            SIL-20AC / 20ACHT
            Equipped with Cooling Function
            The SIL-20AC is equipped with a sample cooler that incorporates a
            dehumidifying function. Samples can be maintained at a fixed temperature in the
            range of 4°C to 40°C. The high cooling speed makes it possible to keep easily
            decomposed sample constituents in a stable condition.
            Rack Changer II
            Supporting Multi-sample Processing
            Maximum 12 microplates can be set. The robot arm in the rack changer
            automatically loads both micro-plates and racks for standard 1.5 mL vials into the
            autosamplers. It is a powerful tool for the  analysis of a large number of samples.

                Sample Carryover Reduced to an Absolute Minimum

            Adsorption of sample constituents has been reduced to an
                                                                 Easily adsorbed compounds   They are hardly detected at
            absolute minimum by using a special processing technology   can be handled with the SIL-20A.  all with a rinsing pump.
            for the sampling needle (patent pending) and rethinking the
            structure of the needle seal and the materials used in flow-line
            parts. As a result, there is hardly any sample carryover. Also,
            the adoption of a PEEK rotor seal allows use over a wide pH                           Not detected
            range, from strongly acidic conditions to strongly basic
            conditions. Using the optional rinse kit (228-43042-91)
            makes it possible to rinse the sampling needle with two
            different solvents, selected in accordance with the purpose.  Cross-contamination Test for Chlorhexidine

                High Throughput

            Only 10 seconds* are required for sample injection.
            High-speed vertical motion of the needle enables
            ultra-high-speed sample processing, which was considered
            impossible with conventional autosamplers. Using in
            combination with a high-speed separation column makes an
            analysis cycle of less than one minute a reality.
            *Specified condition

                Precise Sample Injection
            Greater accuracy has been attained by incorporating a   Injection-Volume Accuracy  Injection-Volume Precision
            high-performance sampling device that can measure out the   Set value (μL)  Measured value (μL)  Injection volume (μL) Area reproducibility (%RSD)
            samples with high precision. The design reflects an emphasis   1  0.99           1           0.43
                                                                  2          1.99            2           0.25
            on basic performance as well as functionality. Also, using
                                                                  5          5.01            5           0.06
            direct injection means valuable samples are not wasted.
                                                                 10         10.00            10          0.04
                                                                 20         19.92            20          0.03
                                                                 50         49.90            50          0.10
                                                                 100        99.70           100          0.11