Page 8 - Shimadzu PDA-8000
P. 8

Shimadzu's Unique Pulse Distribution Analysis Photometry





            Pulse Height Distribution Analysis (PDA) Method Provided as Standard

            The spectral intensity signal from the detector obtained for each pulse discharge is fed in for each analysis element, and
            converted into a frequency distribution. By performing statistical processing appropriate for each analysis element, the
            measurement accuracy can be improved and information on the element form can be extracted.





                       Discharge crater









                       Light emitting
                       section



                                      1 discharge               6 discharges             2000 discharges


                        C        Spectral intensity                                          ・・・・・・・・






                        Si                                                                   ・・・・・・・・





                        Mn                                                                   ・・・・・・・・






            Internal Standard Element Monitoring

            The spectral intensity of internal standard elements is monitored, and   Spectral intensity   IH
            only pulses with the intensity within the specified range are processed                       Valid range
            as valid. Increased accuracy is achieved by eliminating the data outside                      IL
            the range from the statistical processing.

                                                                                  Internal standard element  Time
                                                                         IH  IL :  Discriminator level
                                                                           Valid data
                                                                           Invalid data








      8
   3   4   5   6   7   8   9   10   11   12   13