Page 9 - Shimadzu ICPE-9800 Series
P. 9
Three Features Reduce Gas Consumption by Half
Gas Usage During Analysis and Standby Halved
Mini Torch
Mini-torch is designed to have only half the cross sectional area of standard torches. Applying the same
high-frequency output to the smaller cross section increases the energy density per cross section. This improves the
MINI excitation efficiency, resulting in higher sensitivity.
Eco Mode
20
Reducing the high-frequency output to 0.5 kW
and the plasma gas flow rate to about 5 L/min
during standby saves energy and reduces costs. 10
The analysis mode starts up smoothly as well,
which enables stable analysis.
0
Standard Torch Mini Torch Eco Mode
(conventional) (with mini torch)
Reduced Startup Gas Consumption Vacuum
To enable measurements of wavelengths in the vacuum UV region (below 190 nm) ICPE-9800 systems employ a
vacuum pump (rotary pump) to evacuate the air and remove oxygen from inside the spectrometer. This ensures a
u
u
u
u
uu
u
uum
uu
uum
u
u
u
u
u
u
uu
u
u
ac
ac
cu
cu
cu
Vacu
Vacuumuum
Vacuum
Vacu
Vacu
Vacu
u
u
u
uum
u
u
cu
cu
cu
u
u
shorter startup time in comparison to purge-type systems. When the pump stops, a solenoid valve automatically
closes to maintain the vacuum inside the spectrometer. Consequently, there is no need to worry about
contamination from atmospheric air flowing back into the spectrometer after the pump stops.
ICPE-9800
Advantages of Vacuum Spectrometers 100
Fast startup!
80
Lower operating costs Purge-type
system
Faster startup and shutdown Spectral Intensity 60
Startup takes about 10 minutes. Also, the system can 40
be switched OFF immediately after measurements. 20
0
0 50 100 150 200
Elapsed Time (min)
Change in Spectral Intensity
Sample Introduction System Achieves Stable Analysis for Natural Aspiration and Gravity Drain
Long Periods of Time
The sample introduction system incorporates a high-accuracy nebulizer for introduction
by free aspiration and a gravity drain. Therefore, a peristaltic pump is not necessary.
Since there is no need to worry about a reduction in the efficiency or the variability of
introduction caused by the deterioration of the pump tube, or trouble caused by
improper draining, stable analysis can be accomplished for long periods of time. (If you
wish to automatically add an internal standard element, an optional peristaltic pump
should be used.)
ICPE-9800 Series
Simultaneous ICP Atomic Emission Spectrometers 9