Smart Performance

                     Provides Higher Sensitivity and Reduces Operational Costs


                     A large-capacity differential exhaust turbomolecular pump and a high-performance flow controller ensure that the      New Flow Controller Achieves Exceptional Reproducibility                        Split filter
                     utmost in sensitivity is obtained under all conditions used for GC. This sensitivity combined with high-speed analysis
                     capabilities helps maximize laboratory efficiency by reducing analysis times. In addition, the system can be safely   A new flow controller (AFC) with a CPU uses various control methods to control carrier
                     operated using carrier gases other than helium, such as hydrogen or nitrogen, thereby reducing operating costs.    gas flow to a constant flow speed, flowrate, or pressure. It can also accurately trace the
                                                                                                                                        analytical conditions already being used.
                     Technology Achieves High Sensitivity                                                                               The split line filter can be replaced without any tools. Internal contamination can be
                                                                                                                                        confirmed visually, ensuring filters are replaced at the proper time.

                       Ion Source Featuring High Sensitivity and Long-Term Stability
                     The filament and the ion source box have been separated, which reduces
                     the impact of the filament potential on the interior of the ion source.
                     In addition, a shield is provided to block radiant heat generated by the
                     filament, achieving a uniform temperature inside the ion source box. Thus,                                                                                                                          Valve
                     active spots inside the ion source are not prone to occur, enabling                                                                                                                            Flow Controller
                     high-sensitivity analysis with long-term stability. (Patent: US7939810)                                                                                                                         (AFC-2030)

                                   Shield  Temperature
                                           Low           High
                                              : Filament
                                              : Electric field
                                              : Heat rays
                                                                                                                                                                                           One Touch Inlet Maintenance
                                  Shield
                                                                                                                                                                                        The injection port can be opened or closed without tools
                                                                                                                                                                                                            ™
                       (×1,000)                 (×1,000)                                                                                                                                by simply sliding the ClickTek  lever. Replace the insert,
                     1.00  274.00             3.0  354.00
                        246.00                   306.00                                                                                                                                 slide the lever and feel the click for a leak-free install every
                                              2.5
                     0.75                                                                                                                                                               time.
                                              2.0
                     0.50                     1.5
                                              1.0
                     0.25
                                              0.5
                         14.25  14.50  14.75  15.00  15.25  15.50  12.00  12.25  12.50  12.75  13.00  13.25
                          Phenthoate (5 ng/mL)      Dithiopyr (5 ng/mL)
                       Large-Capacity Differential Exhaust System
                     Through adoption of a new type of turbomolecular pump with increased exhaust efficiency, the
                     system's performance is greatly improved when using hydrogen or nitrogen, in addition to                                                                                            ClickTek Nut
                     helium, as the carrier gas. Moreover, a differential exhaust method is used to create a vacuum
                     separately for the ion source and the quadrupole. As a result, an optimal MS state can be
                     realized regardless of the conditions imposed by the carrier gas used.


                       High-Speed Scan Control Technology
                       Advanced Scanning Speed Protocol (ASSP )                                                                                                                                                    Area value Retention time
                                                                          ™
                                                                                                                                                                                                                    %RSD
                                                                                                                                                                                                                           %RSD
                                                                                                                                                                                                       Acenaphthylene  0.969  0.005
                     The rod bias voltage is automatically optimized during                                                                Advanced GC Oven                                            Fluorene     0.918  0.007
                                                                                                                                                                                                                           0.006
                                                                                                                                                                                                                    1.075
                                                                                                                                                                                                       Phenanthrene
                     high-speed data acquisition, which minimizes                                      Diazinon                                                                                        Anthracene   1.141  0.007
                                                                                                                                                                                                                    1.263
                     sensitivity deterioration during high-speed scans of             Propyzamide                                       The improved temperature control                               Pyrene       1.405  0.004
                                                                                                                                                                                                       Benz[a]anthracene
                                                                                                                                                                                                                           0.005
                     10,000 u/sec or faster. The sensitivity obtained is at                                                             function enables more precise                                  Chrysene     1.283  0.005
                     least five times better than with conventional systems.                                                            temperature control of the GC oven,                            Benzo[b]fluoranthene  1.940  0.003
                                                                                                                                                                                                                    1.268
                                                                                                                                                                                                                           0.003
                                                                                                                                                                                                       Benzo[k]fluoranthene
                     This is effective for scan data sensitivity improvements                                                           which improves the precision of                                Benzo[a]pyrene  0.781  0.005
                     and favorable mass spectrum acquisition, particularly            Black: 1,111 u/sec                                retention time reproducibility.                                Indeno[1,2,3-cd]pyrene 0.744  0.004
                                                                                                                                                                                                                           0.004
                                                                                                                                                                                                       Dibenz[a,h]anthracene 0.836
                                                                                      Red   : 5,000 u/sec
                     in high-speed analysis with Fast-GC/MS, simultaneous             Blue : 10,000 u/sec                               In addition, three oven cooling rate   17.45  17.50  17.55     Benzo[ghi]perylene  0.767  0.004
                     Scan/SIM, FASST analysis, and applications using GC ×   Newly Patented Technology (ASSP)                           levels can be specified to minimize   Mass Chromatogram of Benzo[a]pyrene  Repeatability with Polycyclic
                     GC-MS. (Patent: US6610979)                                                                                         damage to column liquid phases and   (Overlaid plotting of measurements repeated eight times)  Aromatic Hydrocarbons (PAHs)
                                                                                                                                        maximize the service life.