Page 1 - Shimadzu AIM-9000
P. 1

C103-E103E
 Specifications  AIM-9000
 Hardware
 Measurement Mode  Transmission/reflection/ATR  Options  Wide-field camera
 Optical System  15× Cassegrain objective  (independent illumination from four directions)
 15× Cassegrain condenser  ATR objective (Ge)
 MCT Detector  Wavenumber range: 5,000 to 700 cm -1  (narrow band)  ATR pressure sensor
    5,000 to 650 cm -1  (middle band)  Grazing angle objective (GAO)
 Liquid nitrogen monitoring function  Micro vise holder
 TGS Detector  Wavenumber range: 4,600 to 400 cm -1  Infrared polarizer
 (Option)  Automatically switches between detectors, if equipped with   Visible polarizer
 multiple detectors  Installation Conditions  Environmental requirements for performance warranty:   Infrared Microscope
 Supports for   Auto Focus  15°C to 30°C with max. 70%RH (and no condensation)
 Observation and   Automatic adjustment function for Condenser mirror  Site requirements:
 Measurements  Automatic ATR measurement (with purchase of pressure sensor)  10°C to 35°C with max. 70%RH   AIM-9000
 (and no condensation)
 Or max. 60%RH when 30°C is exceeded
 Power Voltage  100/120/220/230/240 V AC
 Power Requirements  85 VA

 Software
 Computer Requirements  OS: Microsoft ®  Windows ®  7 Professional 64bit  Data Processing  • Baseline correction (zero/3-point/multi-point)
       Microsoft ®  Windows ®  10 Pro 64bit  • Advanced ATR correction
 RAM: 8 GB or more  • Atmospheric correction
 Monitor resolution: Min. 1920 × 1080 pixels  • Kramers-Kronig analysis
 Measurement Support  Automatic aperture setting   • Differential analysis
 Functions  (automatic contaminant recognition system)  • Mapping chemicals based on calculation formula
 Zoom function (max. 330× digital zoom)   • Mapping chemicals based on principal component analysis
 Multi-image tiling  • Mapping chemicals based on similarity to reference spectra
 Up to 60 aperture settings can be specified on microscope   • Spectral search
 or tiled images.  • Pasting spectra and images into other applications
 Validation  Japanese Pharmacopoeia
 European Pharmacopoeia
 Chinese Pharmacopoeia
 Confirmation of aperture size
 Option  Mapping program


 Examples of Combined Systems
















 ™
 ™
 IRTracer -100 +AIM-9000  IRAffinity -1S+AIM-9000


 This product is certified as Shimadzu's Eco-Products Plus.
 Reduced power consumption by 55% compared with conventional Shimadzu's products.




 LabSolutions,IRTracer, and IRAffinity are trademarks of Shimadzu Corporation.
 Microsoft and Windows are registered trademarks of Microsoft Corporation in the United States and/or other countries.






 For Research Use Only. Not for use in diagnostic procedures.
 This publication may contain references to products that are not available in your country. Please contact us to check the availability of
 these products in your country.
 Company names, products/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation,
 its subsidiaries or its affiliates, whether or not they are used with trademark symbol “TM” or “®”.
 Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether
 or not they are used with trademark symbol “TM” or “®”.
 Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
 www.shimadzu.com/an/  The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice.
 Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
 © Shimadzu Corporation, 2021
 First Edition: May 2016   3655-01112-PDFIT
   1   2   3   4   5   6