Page 50 - Application Handbook - TOC
P. 50

Sum parameter – Total Organic Carbon




                                         Carbon dioxide determination in beer

                   SCA-130-403


            Carbon dioxide is an important ingredient in        In order to develop a method that does not
            many soft drinks. This is also the case for         have these disadvantages, a TOC analyzer
            beer. It creates a sparkling and refreshing         was used.
            (tangy) taste and is important for the
            formation of foam.                                  ■ Innovative methods
                                                                In this method, the sample (beer) is directly
            The CO 2 content of a beer affects the              placed in a 40 mL autosampler vial. 5 mL of a
            threshold values for various fragrance and          32% NaOH solution was added to the
            aroma components. In addition, bottling             autosampler vial to preserve the CO 2.
            under CO 2 increases the shelf life of beer..
                                                                The sample is subsequently added directly to
                                                                the autosampler and the IC  (inorganic
                                                                carbon) content is measured.





















                                                                Preservation step:
            In the manual of the ‘central- European
            brewery technological analysis commission)’
            (MEBAK)      various    methods     for   the                            -             -
            determination of CO 2 are listed.  These are                   CO  + OH     →   HCO
                                                                                                  3
                                                                              2
            generally based on manometric or titrimetric
            method, or they are methods that use                                   -           2-
            specialized detectors.                                    CO  + 2 OH      →  CO  + H O
                                                                         2                    3      2
            Disadvantages of these methods are often
            the lack of selectivity for CO 2 (other gases or    In the TOC analyzer, the sample is injected in
            substances are also determined), high               a concentrated phosphoric acid solution
            expenditure in terms of personnel and time,         (25%). The CO 2 is subsequently released
            and the lack of possibilities for automation.       again and is transferred via the carrier gas to
                                                                a CO 2-selective NIDR detector where it is
                                                                detected.