Page 17 - 5_TOC in daily practice
P. 17

SCA-130-508



            ■ Instrument blank value                            ■ Reagents – blank value
            Due to the ubiquity of certain compounds,           Reagents,  such  as  hydrochloric  acid,  often
            potential  minute  leaks  in  the  instrument  and   stand unobserved and unintentionally next to
            possibly persistent deposits or biofilms, can       the instrument for months and absorb vast
            lead to area values that originate from the         amounts  of  organic  carbon  from  the
            instrument  itself  and  not  from  the  measured   environment     (‘from   a    blank    value
            sample. To determine  this value, the ‘blank        perspective’). For the determination of such
            check’ procedure can be performed:                  reagent blank values, the standard addition
                                                                method is suitable.
            For this purpose, the  system (TOC-LCPH)
            carries  out  an  automatic  analysis  of           ■ Contaminations from the environment
            circulating  ultrapure  water.  The  resulting      At last, the blank value is discussed that can
            condensation is collected in a suitable             arise  from  everywhere  in  the  environment
            container inside the  instrument. As soon  as       and  can  enter  the  analysis  from  various
            enough condensed water is available, the            sources.    People    and   their   industrial
            water is circulated,  i.e. it is injected again.    landscapes are the source of a large variety
            This procedure is carried out 50 times and it       of organic carbon compounds. Humans
            can, therefore, be assumed that the final           themselves consist of 18.2 % organic carbon
            determined area value corresponds to the            and lose, for example, 1 – 2 g skin particles
            actual instrument blank value.                      per day. These generally settle in the form of
                                                                house dust.

                                                                 In addition to carbon originating from our
                                                                own bodies, carbon sources present in
                                                                cosmetics  or  toiletries  such  as  soaps,
                                                                deodorants,  perfumes,  after-shaves  skin
                                                                creams, ointments, plays an important role.
                                                                Also the laboratory harbors large sources of
                                                                organic carbon compounds.

                                                                Room air contains numerous pollutants that

                                                                can easily  bind to dust particles and can
                                                                therefore also be present in house dust.
            Note:  The  determination  of  the  instrument      Finally, it should not be overlooked that room
            blank  value  is  also  suitable  for  intensive    air  contains  approximately  0.4  vol%  /  0.06
            cleaning of the flow lines and for the ‘breaking    mass% (400 / 600 ppm) CO 2.
            in’ a newly built-in catalyst. As this is a very
            time    consuming     method,     it   should
            preferentially be carried out overnight.