Page 8 - Solution SystemGC for Hydrocarbon
P. 8

Fast Natural Gas Analyzer

            Nexis GC-2030FNGA1 with Hydrogen / Nexis GC-2030FNGA2 without Hydrogen




              Analyzer Description                     Typical Chromatogram


              System Configuration:
              ・ 4 valves / 8 columns (packed and
                 capillary) / 2 TCDs / FID
              Sample Information:
              ・ Natural gas or similar gaseous mixture
              Compounds Analysed:
              ・ C1 to C5 (Methane, Ethane, Propane,
                 i-Butane, n-Butane, i-Pentane,
                 n-Pentane)
              ・ C6+ as Back  ush

              ・ He, H 2, O 2, N 2, H 2S, CO, CO 2
              Typical  Concentration Range:
              ・ 0.01% mol for permanent gases and
                 C2 hydrocarbon on TCD, 10 ppm
                 for C3 to C6+ hydrocarbons on FID
                 except for H 2S and CH 4, 0.10% mol
                 for H 2S and 20% mol for CH 4
              Reference Method:
              ・ ASTM D-1945, D-3588 and GPA2261
    Natural Gas


              Features


              ・  Dual TCD channels
              ・  Dedicated channel for Hydrogen
                 and Helium analysis
              ・  FID channel for C3 to C6+
                 hydrocarbons
              ・  Rugged packed and capillary
                 columns
              ・  10 mins analysis time
              ・  BTU calori  c calculation software
                 provided

















      8
   3   4   5   6   7   8   9   10   11   12   13