Page 15 - Shimadzu TOC-L
P. 15
Specifications
˙ TOC-L Series Total Organic Carbon Analyzer
*UFNT )JHI 4FOTJUJWJUZ .PEFM 4UBOEBSE .PEFM
Model TOC-LCPH TOC-LCSH TOC-LCPN TOC-LCSN
Measurement Method 680 °C combustion catalytic oxidation – non-dispersive infrared detection (NDIR) method
Operation Method PC-controlled Standalone *PC control available PC-controlled Standalone *PC control available
Measured Items TC, IC, TOC (= TC-IC), NPOC (TOC measurement via acidification and sparging) *Option: POC, TOC (= NPOC + POC), TN
Applicable Samples Aqueous (optional solid/gas samples)
TC: 0 to 30,000 mg/L TC: 0 to 30,000 mg/L
IC: 0 to 35,000 mg/L IC: 0 to 3,000 mg/L
Measurement
Range (Option) (Option)
TN: 0 to 10,000 mg/L TN: 0 to 10,000 mg/L
POC: 0 to 500 mg/L POC: 0 to 500 mg/L
Detection Limit TC, IC: 4 µg/L, TN: 5 µg/L TC: 50 µg/L, IC: 4 µg/L, TN: 20 µg/L
TC, IC, NPOC: CV 1.5 % max. or ±4 µg/L max TC, NPOC: CV 1.5 % max. or ±50 µg/L max, IC: CV 1.5 % max. or ±4 µg/L max
Reproducibility
(Optional TN: CV 3.0 % max. or ±5 µg/L max) (Optional TN: CV 3.0 % max. or ±20 µg/L max)
Measuring Time TC: approx. 3 min, IC: approx. 3 min (Optional TN: approx. 4 min) TC: approx. 3 min, IC: approx. 4 min (Optional TN: approx. 4 min)
Sample Injection Automatic sample injection using a syringe pump and slide type injection mechanism
Sample Injection 10 to 2,000 µL variable TC: 10 to 150 µL variable, IC: 10 to 4,500 µL variable
Volume
IC Removal Automatic addition of acid and sparging
Sample Dilution Dilution rate of 2× to 50× (automatic sample dilution by syringe pump), dilution accuracy: ±2 % max. (2× to 20×), ±5 % max. (21× to 50×)
Operation by color LCD screen Operation by color LCD screen
Display and and keyboard and keyboard
Operations Operated by PC Operated by PC
*Operation by PC is also possible *Operation by PC is also possible
External Memory — USB flash memory used — USB flash memory used
(Standalone Type)
Printer — Portable thermal printer and — Portable thermal printer and
(Standalone Type) PC USB printer can be used PC USB printer can be used
High-purity air (CO, CO2, HC content: Each 1 ppm max., dew point: -50 °C max.)
Carrier Gas Supply pressure: 200±10 kPa (Additional use of optional carrier gas regulator: 300 to 600 kPa)
Optional use of nitrogen gas (not possible in the TN measurement). With the standard model, optional use of pressurized gas.
150 mL/min (230 to 250 mL/min during sparging) 230 mL/min (A separate 100 mL/min is required for
Gas Consumption
(Variable flow rate) sparging with ASI-L. (variable flow rate))
Power Supply 100 to 240 V AC, 600 VA (Permitted range: 90 to 264 V AC)
Applicable Regulations CE
Ambient
Temperature Range 5 to 35 °C
Dimensions W340 × D660 × H480 mm (Excluding protrusions. For details, see the External Dimensions Diagram.)
Weight Approx. 35 kg
˙ ASI-L Autosampler ˙ TNM-L TN (Total Nitrogen) Unit ˙ SSM-5000A Solid Sample Combustion Unit
Select from three types: Measurement TC Oxidation Combustion catalytic oxidation
Vial Types Chemiluminescence
9 mL, 24 mL, 40 mL Method Method (TC furnace temperature: 900 °C)
Number of Vials 9 mL: 93, 24 mL: 93, 40 mL: 68 Measured Items TN (total nitrogen) IC Reaction Acidification
With dedicated septum Measurement Method (IC furnace temperature: 200 °C)
Vial Septum 0 to 10,000 mg/L
(excluding 9 mL vials) Range Measured Items TC, IC, TOC
Possible 5 µg/L (CPH, CSH) TC: 0.1 to 30 mg carbon
Sample Sparging (The optional external sparging Detection Limit Measurement TC: (1 to 20 µg carbon in
kit is required.) 20 µg/L (CPN, CSN) Range high-sensitivity measurement)
W370 × D540 × H490 mm Reproducibility CV 3 % max. IC: 0.1 to 20 mg carbon
Dimensions
(excluding protrusions) Measuring Time Approx. 4 min Maximum 1 g
Sample Amount (aqueous content: < 0.5 g)
Weight Approx. 14 kg Air
Ozone Source Gas (compressed air or housing air) Measuring Time Normally 5 to 6 minutes
500 mL/min
˙ OCT-L 8-Port Sampler W270 × D240 × H160 mm Carrier Gas 99.9 % O2 at 500 mL/min
High-purity O2 gas is required for
Dimensions high-sensitivity measurement.
Number of OCT-L Up to 2 OCT-L units (excluding protrusions)
Units Connected for a single TOC-L Weight Approx. 6 kg 100 to 127 or 220 to 240 V AC
Power Requirements
Vial Types Any sample container can be used as ordered, 700 VA
8 for a single OCT-L Dimensions W450 × D656 × H290 mm
Number of Vials
16 for dual OCT-L ˙ Recommended PC Speci cations Weight Approx. 30 kg
No sparging with OCT-L; sparging
Sample Sparging
is done in the TOC-L syringe
W245 × D245 × H440 mm
Dimensions
(excluding protrusions)
Weight Approx. 3.5 kg