Page 3 - Shimadzu Nexera lite inert
P. 3

Improved Data Reliability by Preventing System Corrosion by Halogenic Salt
 ੜମߴ෼ࢠͷ෼ੳʹ࠷ద
 Nexera lite inert  Aggregates of antibody drugs show immunogenicity and cause a decrease in puri cation ef ciency and production.

               Therefore, an aggregation test is one of the most critical evaluations when manufacturing biopharmaceuticals and
               controlling their quality. However, a metal-free LC system is required for good repeatability over a long time due to
               high-concentration, salt-containing mobile phases such as Sodium chloride. With no metal material in the  ow path, Nexera
               lite inert provides stable aggregate assay results with good repeatability.
                  mAU
 The analysis of proteins such as antibody drugs requires careful operation because a high-salt mobile
               800                     monomer
 phase sometimes corrodes the system. In addition, the adsorption of proteins to metal surfaces often   mAU  Analytical Conditions
                    6 aggregates                              Mobile phase   0.2 mol/L  Potassium phosphate buffer
 results in poor chromatographic performance.   4                        containing  0.25 mol/L Potassium chloride (pH 6.2ʣ
 Nexera lite inert is a robust HPLC system that eliminates the risk of corrosion by high-salt mobile phases   600  2  Flow rate    0.5 mL/min
                    0
 and sample adsorption onto metal surfaces. It improves the data quality of protein analysis in biological   -2  fragments  Column    TSKgel G3000SWXL
 samples with superior reproducibility, without any special considerations for stable and long-term use.  -4  10.0  15.0  20.0     (300 mm×7.8 mmI.D., 5 µm)
               400              min                           Column temp.  Ambient
                                                              Detection    UV, 280 nm

               200
                                                              Rusting

                0
                 0.0   5.0    10.0  15.0   20.0  25.0  min        Stainless-steel pump   LC-40i
                       Monoclonal IgG antibody aggregate analysis  Corrosion resistance test results of pump heads
                          by size exclusion chromatography




               Superior Retention Time Stability by Highly Accurate Solvent Delivery

               Thanks to the outstanding  ow-rate stability, the Nexera lite inert provides excellent retention time repeatability.
                                                              Analytical Conditions
                                                              Mobile phase:  0.1 mol/L Potassium phosphate buffer
                                 Repeatability of retention time:
                                                                         (pH 7.0) containing 0.2 mol/LSodium
                                 % RSD =0.08
                                                                         chloride aq.
                                                              Flow Rate:   0.5 mL/min
                                                              Column:    Shim-pack Bio Diol
                                                                         (300 mm×4.6 mm I.D., 5 µm)
                                                5  injection
                                                th
                                                              Column temp.: 25ˆ
                                                th
                                               ̐  injection   Detection:   UV, 280 nm
                                                3  injection
                                                rd
                                               2  injection
                                                nd
                                                1  injection
                                                 st
                 0.0       5.0      10.0      15.0     min
                    Analysis of the aggregates in the antibody drug (n=5)



               Real-Time Monitoring of Mobile Phase pH


               In ion exchange or size exclusion chromatography, the pH of the mobile phase sometimes affects
               the separation of compounds. The pH Monitor, pHM-40, continuously monitors and records the
               pH of the mobile phase. The pH and detector chromatograms are saved within the same data  le
               and can be overlapped, ensuring ef cient data traceability.



                                                                                          Nexera lite inert
                                                                                             High Performance Liquid Chromatograph
   1   2   3   4   5   6   7   8