Page 4 - Shimadzu i-Series Method Transfer System
P. 4

Dual Flow Lines Simplify the Process

                     of Migrating Methods



                     A single integrated system allows analyses in two flow lines with UHPLC and HPLC delay volumes. In addition to
                     simplifying the transfer of analytical methods for customers using HPLC, it can also streamline the process of
                     converting customers’ analytical methods from HPLC to higher-speed UHPLC.



                       Supports Method Compatibility Compliant with

                       the United States Pharmacopeia


                     General Chapter <621> Chromatography in the United States Pharmacopeia specifies allowable ranges for changes to
                     analytical parameters, so that revalidation is not necessary as long as the setting values remain within those allowable
                     ranges. i-Series MT system includes dual flow lines for HPLC and UHPLC analysis. With two system volumes, it can
                     support both existing analytical methods and UHPLC methods.





                     Flow line 1
                     HPLC Conditions
                                                          mAU
                     Column     : Shim-pack™ VP-ODS    2.25
                                 150 mm L. × 4.6 mm I.D., 5.0 µm
                                                       2.00
                     Mobile phase A   :  Phosphate (potassium) buffer
                                 solution, pH 3.5      1.75                            Ketoprofen (82.43%)
                     Mobile phase B   :  Acetonitrile  1.50                                       Impurity C (9.98%)
                     Composition   :  A/B = 57/43 (v/v)  1.25                              Impurity B (5.22%)
                     Flow rate   :  1.0 mL/min         1.00               Impurity A (2.37%)
                     Column temperature :  40°C        0.75
                     Injection volume   :  20 µL       0.50
                     Sample     :  Ketoprofen 1 mg/mL  0.25
                                                       0.00
                                                         0.0  1.0  2.0  3.0  4.0  5.0  6.0  7.0  8.0  9.0  10.0  11.0  12.0  min




                     Flow line 2
                     UHPLC Conditions                   10  mAU
                     Column     :  Shim-pack XR-ODS II   9
                                 75 mm L. × 3.0 mm I.D., 2.2 µm  8
                     Mobile phase A   :  Phosphate (potassium) buffer   7
                                 solution, pH 3.5        6                             Ketoprofen (81.71%)
                     Mobile phase B   :  Acetonitrile                                             Impurity C (10.24%)
                                                         5
                     Composition   :  A/B = 57/43 (v/v)
                                                         4                                Impurity B (5.60%)
                     Flow rate   :  1.0 mL/min                            Impurity A (2.44%)
                                                         3
                     Column temperature :  40°C
                                                         2
                     Injection volume   :  20 µL
                                                         1
                     Sample     :  Ketoprofen 1 mg/mL
                                                         0
                                                         0.00  0.25  0.50  0.75  1.00  1.25  1.50  1.75  2.00  2.25  2.50  2.75  min










        4
   1   2   3   4   5   6   7   8