Page 5 - Shimadzu GC-2014
P. 5

High Performance
 Performance lifted to a higher level

 The Highest available precision and accuracy similar to our industry leading GC-2010



 Digital carrier gas control  AFC  Our New FPD is used for all columns
 Single or Dual AFC  exibility  Enables higher-level analyses  Detectors


 Higher-level repeatability of carrier gas is indispensable   This digital control is standard for not only capillary   The detectors have been completely redesigned,   used in TCD-14. The newly designed FPD-2014 takes
 to high data reliability. More accurate electronic  ow   columns, but packed columns as well.   incorporating the GC-2010 detector designs for capillary   advantage of the holophotal construction of the
 controllers set and maintain  ow rates in multiple modes   Accurate  ow rate control via AFC has higher-level   analyses and the GC-14 detector designs for packed   FPD-2010. This is the simplest nozzle replacement no
 automatically eliminating human error.  repeatability of retention time and peak area, enabling a   columns. This TCD-2014 unit is ideal for packed column   matter if you are using capillary or packed columns.
 The GC-2014 Series is equipped with the advanced  ow   higher level of analyses.  measurements employing the semi-diffusion cell design
 controller (AFC) technology inherited from the GC-17A
 Series and GC-2010 Series.   Holophotal Flame Photometric Detector  Quartz glass nozzle (for packed column analysis)





 Dodecane  Tetradecane  Hexadecane
 RT            Area  Height  RT            Area  Height  RT            Area  Height  Stainless steel nozzle (for capillary column analysis)
 1  5.243  55397  34356  1  6.731  55379  33635  1  8.06  55898  33437
 2  5.243  55418  34431  2  6.731  55529  34446  2  8.06  56170  33499
 3  5.243  55762  34571  3  6.731  55880  34042  3  8.059  56486  33610
 4  5.243  55632  34497  4  6.731  55717  34551  4  8.06  56347  32899  Nozzle  Lens   Filter  Photomultiplier
 5  5.244  55861  34865  5  6.732  56021  34611  5  8.061  56572  33931
 6  5.243  55957  35396  6  6.731  56060  34677  6  8.06  56780  33086
 7  5.243  56026  35066  7  6.731  56120  34426  7  8.061  56714  34300
 8  5.242  56083  35439  8  6.73  56164  34638  8  8.059  56694  33358
 9  5.243  55770  34739  9  6.73  55937  33984  9  8.059  56509  34382
 10  5.243  55857  34614  10  6.731  55762  34062  10  8.06  56510  33427
 Average (Avg.)  5.243  55776.3  34797.4  Average (Avg.)  6.7309  55856.9  34307.2  Average (Avg.)  8.0599  56468  33592.9
 Standard Deviation  0.000471  234.7737  387.9167  Standard Deviation  0.000568  258.8747  352.4608  Standard Deviation  0.000738  269.1274  482.072
 C.V.%  0.008991  0.42092  1.114786  C.V.%  0.008433  0.463461  1.027367  C.V.%  0.009155  0.476602  1.435041
                                                Simple nozzle replacement supports both
                                                capillary and packed columns.








                                                               Analytical capacity is increased
            Unsurpassed accuracy                               while maintaining small footprint
            Injection Units                                    Column Oven


            The design of the SPL-2014 capillary column sample   By using the GC-2010 control electronics and cooling
            injection unit is based on the GC-2010 technology.  mechanisms, column oven performance has been greatly
            This accuracy was unattainable with previous models.  improved over that of the GC-14 Series.
            The packed column sample injection unit employs the   Oven capacity is increased while keeping the same width
            proven design of the GC-14 injection unit.         as the GC-14, with enough capacity to accommodate
                                                               both capillary and packed columns together.




                                                                                                     GC-2014
 4                                                                                           Capillary and Packed Gas Chromatograph  5
   1   2   3   4   5   6   7   8   9   10