Page 5 - Shimadzu GC-2014
P. 5
High Performance
Performance lifted to a higher level
The Highest available precision and accuracy similar to our industry leading GC-2010
Digital carrier gas control AFC Our New FPD is used for all columns
Single or Dual AFC exibility Enables higher-level analyses Detectors
Higher-level repeatability of carrier gas is indispensable This digital control is standard for not only capillary The detectors have been completely redesigned, used in TCD-14. The newly designed FPD-2014 takes
to high data reliability. More accurate electronic ow columns, but packed columns as well. incorporating the GC-2010 detector designs for capillary advantage of the holophotal construction of the
controllers set and maintain ow rates in multiple modes Accurate ow rate control via AFC has higher-level analyses and the GC-14 detector designs for packed FPD-2010. This is the simplest nozzle replacement no
automatically eliminating human error. repeatability of retention time and peak area, enabling a columns. This TCD-2014 unit is ideal for packed column matter if you are using capillary or packed columns.
The GC-2014 Series is equipped with the advanced ow higher level of analyses. measurements employing the semi-diffusion cell design
controller (AFC) technology inherited from the GC-17A
Series and GC-2010 Series. Holophotal Flame Photometric Detector Quartz glass nozzle (for packed column analysis)
Dodecane Tetradecane Hexadecane
RT Area Height RT Area Height RT Area Height Stainless steel nozzle (for capillary column analysis)
1 5.243 55397 34356 1 6.731 55379 33635 1 8.06 55898 33437
2 5.243 55418 34431 2 6.731 55529 34446 2 8.06 56170 33499
3 5.243 55762 34571 3 6.731 55880 34042 3 8.059 56486 33610
4 5.243 55632 34497 4 6.731 55717 34551 4 8.06 56347 32899 Nozzle Lens Filter Photomultiplier
5 5.244 55861 34865 5 6.732 56021 34611 5 8.061 56572 33931
6 5.243 55957 35396 6 6.731 56060 34677 6 8.06 56780 33086
7 5.243 56026 35066 7 6.731 56120 34426 7 8.061 56714 34300
8 5.242 56083 35439 8 6.73 56164 34638 8 8.059 56694 33358
9 5.243 55770 34739 9 6.73 55937 33984 9 8.059 56509 34382
10 5.243 55857 34614 10 6.731 55762 34062 10 8.06 56510 33427
Average (Avg.) 5.243 55776.3 34797.4 Average (Avg.) 6.7309 55856.9 34307.2 Average (Avg.) 8.0599 56468 33592.9
Standard Deviation 0.000471 234.7737 387.9167 Standard Deviation 0.000568 258.8747 352.4608 Standard Deviation 0.000738 269.1274 482.072
C.V.% 0.008991 0.42092 1.114786 C.V.% 0.008433 0.463461 1.027367 C.V.% 0.009155 0.476602 1.435041
Simple nozzle replacement supports both
capillary and packed columns.
Analytical capacity is increased
Unsurpassed accuracy while maintaining small footprint
Injection Units Column Oven
The design of the SPL-2014 capillary column sample By using the GC-2010 control electronics and cooling
injection unit is based on the GC-2010 technology. mechanisms, column oven performance has been greatly
This accuracy was unattainable with previous models. improved over that of the GC-14 Series.
The packed column sample injection unit employs the Oven capacity is increased while keeping the same width
proven design of the GC-14 injection unit. as the GC-14, with enough capacity to accommodate
both capillary and packed columns together.
GC-2014
4 Capillary and Packed Gas Chromatograph 5