Page 13 - Shimadzu GC-2014
P. 13

Flexibilty  Application Systems
 Effective Use of GC Series Column Resources  GC-2014 Analysis Systems for Every Application  Speci cations

 Effective use of valuable column resources is maintained by using the 40-mm-pitch   Headspace Analysis System
 glass column the same as Shimadzu GC-7, 9, 12, 14, 15 and 16 Series.
 • High reproducibility ensure reliable quantitation for volatile component   Column Oven  Detectors
 analysis.   Temperature range  (Ambient + 10°C) ~ 400°C (using liquid CO2 gas*: -50°C ~400°C)  Temperature range  400°C max. (FID, TCD, ECD, FTD)  350°C max. (FPD)
 Large Column Oven Facilitates  System Con guration  Dimensions  250 (W) × 360 (H) × 175 (D) mm  Temperature setting  1°C steps
 • GC-2014 + HS-10
                                                                No. of units installed simultaneously
                                                                               Up to 4 units (restricted depending on detector type)
                            15.8L
             Oven capacity
 Column Operations  Analysis Applications  Temperature accuracy  Set value (K) ± 1% (calibration at 0.01°C increments)  Detector type  FID, TCD, ECD, FPD, FTD for capillary/packed
 • Measurement of VOCs in water  Temperature deviation  2°C max. (on 200mm dia. circumference 30mm from rear)  Flame Ionization Detector (FID)
 Column replacement is easier by increasing the   • Blood Alcohol Concentration (BAC)  Temperature variation coef cient  0.01°C/°C  System  Dual  ow rate differential system
 oven capacity while keeping its width the same.  Temperature program steps  Up to 20 (cooling program possible)  Temperature range  400°C max.
             Programmed rate setting range  -250°C ~ 250°C/min  Minimum detected quantity  3pgC/s (dodecane)
 Simulated Distillation GC System for LabSolutions
             Total time for all steps  9999.99 minutes max.     Dynamic range  10 7
 • Measures the boiling point distribution of petroleum distillate using the   Linear heating range  Up to 150°C: 30°C/min (100V/120V), 60°C/min (230V)  Nozzle  Quartz glass Standard: for packed, Option: for capillary
 relationship between retention time and boiling point.  Up to 250°C: 20°C/min (100V/120V), 40°C/min (230V)  Thermal Conductivity Detector (TCD)
 • Analyzes distillation characteristics and prints specialized reports.  Up to 380°C: 10°C/min (100V/120V), 20°C/min (230V)  Dual  ow rate differential system
                            Up to 400°C: 7°C/min (100V/120V), 15°C/min (230V)  System
 GC-2014  System Con guration  *at 25°C ambient temperature     Temperature range  400°C max.
 • GC-2014 + WBI-2014 + LabSolutions GC + Simulated Distillation GC Analysis Software
 250mm  (Select sample injection unit and column according to the target sample)  Cooling rate  300°C ~ 50°C in 6 min max.  (at 25°C ambient temperature)  Dynamic range  10 5
             Columns accepted  Capillary columns: 2             Sensitivity    40,000mV · mL/mg (built-in pre-ampli er, with 10 × ampli cation)
 GC-14  Analysis Applications  Packed columns for GC14B: 4 (Glass columns: 2)  Electron Capture Detector (ECD)
 • Petroleum distillate
                                                                                            63
 230mm       *Optional parts are required to use liquid CO2 gas.  System       Fixed current system using  Ni370MBq radiation source
                                                                Temperature range  400°C max.
 PONA Analysis System  Sample Injection Unit                    Minimum detected quantity  0.1pg/s (γ-BHC)
 • Separates gasoline or other hydrocarbon compounds, identi es the peaks   Temperature range  Up to 400°C  Dynamic range  10 4
 and classi es them by carbon number, or by type (paraf n, ole n,   Heating settings  1°C steps  Flame Photometric Detector (FPD)
 Packed Column  Capillary Column  naphthene and aromatic series) for quantitation.  No. of units installed simultaneously  Up to 3 units  Temperature range  350°C max.
 Accommodates both packed and capillary columns.  System Con guration  Sample injection unit types  Dual packed, single packed, split/splitless, direct, direct (AMC)  Dynamic range  P: 10 4
 • GC-2014 + CRG-2014 + GCsolution + PONAsolution                              S: 10 3
 + MS Excel (commercially available spreadsheet software)  Carrier Gas Flow Controller  Minimum detected quantity  P: 0.5pgP/s (tributyl phosphate)
 (Select sample injection unit and column according to the target sample)
             For Packed / Dual                                                 S: 8pgS/s (dodecane thiol)
 Analysis Applications  Flow rate setting range  0 ~ 100mL/min  Flame Thermionic Detector (FTD)
 Select Your Flow Controller Combination  • Categorization of naptha, gasoline and gasoline-based materials by carbon number or type   Programmable steps  7  (Two types, one for capillary and one for packed. The speci cation are the same.)
 and their quantitation.(Also offers calculation of mean speci c gravity, mean molecular
 weight, and octane value.)  Programmed rate setting range  -400 ~ 400mL/min  Temperature range  400°C max.
 Five slots for manual  ow
             Correction function  Maintains column  ow rate during column oven heating  Dynamic range  N: 10 3
 controllers are in the clear case   Workstations  For Capillary Split/Splitless, Direct  P: 10 3
 atop the main unit.   (Split/splitless injection mode)         Minimum detected quantity  N: 0.4pgN/s (azobenzene)
 LabSolutions LC/GC Ver.5
 Five  AFC/APC slots are in the   Pressure setting range  0 ~ 970kPa           P: 0.05pgP/s (malathion)
 • LabSolutions LC/GC Ver.5 is the next generation of chromatography data system that integrates control of GC* (GC-2010 Plus, GC-2010, GC-2014, GC-2025
 back bottom of the main unit.   and GC-14B) and LC and deliver greater network capability.  Programmable steps  7 (pressure-decreasing program possible)
 * LabSolutions does not support Distillation GC Software and PONAsolution, please use GCsolution for these software.  Programmed rate setting range  -400 ~ 400kPa/min  Display
 Gases are supplied or cut off in
 Manual   Note: CBM-102 is necessary for controlling GC-14B.  Split ratio setting range  0 ~ 9999.9  240 × 320 dot graphics display (30 characters × 16 lines)
 conjunction with switching   controllers  Manual  ow  GCsolution Ver.2  Total  ow rate setting range  0 ~ 1200mL/min
 controllers
 ON/OFF power to the main unit   • GCsolution Ver.2 supports not only GC-2014, but also GC-2025*, GC-2010 Plus*, GC-2010, GC-17A, GC-1700 and GC-14A/B.  Correction function  Maintains column average linear velocity during column oven heating  Dimensions, Weight, Power Requirements (GC main unit)
 * GC-2010 Plus can be controlled by GCsolution Ver.2.32 or later and GC-2025 by GCsolution Ver. 2.40 or later respectively.   (for capillary only)
 or detector even if the low-cost   Note1: CBM-102 is necessary for controlling GC-17A, GC-1700 and GC-14A/B and the data acquisition. CBM-102 is not sold in EU area.  (Pressure mode direct injection)  Dimensions   400 (W) × 690 (H) × 607 (D) mm
 manual controllers are used.   Note2: In controlling GC-14A, the workstation functionalities are partly limited like the detector range 4 cannot be used.  Pressure setting range  0 ~ 970kPa/min  Weight  48kg (GC-2014AF model)
 All parameters for GC analysis   • Both LabSolutions and GCsolution software can control a maximum of four GC units in a computer. By inputting detector analog signal into CBM-102, data   Programmable steps  7   Power Requirements  AC100V/120V 230V
 acquisition is possible even on other Shimadzu GC models that cannot be controlled by the software.    1800VA (GC-2014AF model, AC100V/120V) or
 are controlled digitally when   Programmed rate setting range  -400 ~ 400kPa/min  2600VA (GC-2014AF model, AC230V), 50/60Hz
             Correction function  Maintains column average linear velocity during column oven heating
 APC’s are used to control   Solenoid valve unit  Chromatopac C-R8A *   (for capillary only)
 detector gases. This facilitates   Electronic   (Flow-rate mode direct injection)
 controllers  • Acknowledged data processing functions and ease of operation
 the setting of analytical   • Built-in validation functions  Flow rate setting range  0 ~ 1200mL/min
 conditions and log management   • High-speed RS-232C port provided standard  Programmable steps  7  -400 ~ 400mL/min
             Programmed rate setting range
 by a data station.  • Easy operation designed for factory use  For Single Packed, Direct (AMC)
 * Chromatopac C-R8A is not sold in EU area.
 • C-R8A uses a SD card as the memory device.  Flow rate setting range  0~100mL/min
             Correction function   Maintenance column  ow rate during column oven heating
 AFC

                                                                                                     GC-2014
 11  12                                                                                      Capillary and Packed Gas Chromatograph  13
   8   9   10   11   12   13   14