Page 14 - Shimadzu EDX-LE
P. 14

Instrument Specifications



                     Primary Specifications                              Installation Example                                                   Options
                     Measurement Principle  X-ray fluorescence spectrometry  Dimensions of the Main Unit  W 520 mm × D 650 mm × H 420 mm       Halogen Screening Analysis Kit                     Sample Cells
                     Measurement Method  Energy dispersive              Main Unit Weight   Approx. 60 kg                                      P/N 212-24908-91                                   3571 General Open-End X-Cell (no lid)
                     Measurement Sample Type  Solids, liquids, or powder                                                                      This kit includes an instruction manual for Halogen analysis and a check sample required   P/N 219-85000-55 (100 pcs/set)
                                                                                                                                              for measurement of 6 elements (Cd, Pb, Hg, Cr, Br, and Cl) specified by the RoHS
                     Elements to be Detected  13 Al to  92 U                                                                                  directive and Halogen regulation.                  (Outer diameter: 31.6 mm, volume 10 mL)
                     Sample Chamber Size  Max. W 370 mm × D 320 mm × H 155 mm                                      Unit: mm                                                                      Polyethylene sample cell used for liquid and powder samples. Used
                                                                                                                                                                                                 with Mylar or polypropylene films.
                                                                                              At least 200 mm between unit and wall
                    X-Ray Generator                                                                                                           RoHS, Halogen, and Antimony Screening Analysis Kit  3529 General X-Cell (with lid)
                                                                                                                                              P/N 212-24922-91
                     X-Ray Tube         Rh target
                                                                                                                                              This kit includes an instruction manual and a check sample required for measurement of   P/N 219-85000-52 (100 pcs/set)
                     Tube Voltage       5 kV to 50 kV                                                                                         7 elements including those specified by the RoHS directive, Halogen regulation, and   (Outer diameter: 32 mm, volume 8 mL)
                     Tube Current       1  A to 1,000  A                        650                                                           Antimony (Cd, Pb, Hg, Cr, Br, Cl, and Sb)          Used for liquid samples. Equipped with relief hole and liquid retainer
                     Cooling Method     Air cooling (with fan)                                                                                                                                   in case of liquid expansion.
                     Exposure Area      Automatic switching between 3, 5, and 10 mm                                                           Small Spot Solder Analysis Kit
                                        dia. areas (1 mmø is an option)                                                                       P/N 212-24850-41                                   3577 Micro X-Cell
                     Primary Filter     Automatic switching between: 5 types + OPEN     520        (460)                                      This kit includes an instruction manual for small spot solder analysis and a small spot   P/N 219-85000-54 (100 pcs/set)
                                                                                                                                              collimator plate required for measurement of a print circuit board.  (Outer diameter 31.6 mm, volume 0.5 mL)
                    Detector                                                                                                                                                                     For trace samples. Use with a collimator is recommended to reduce
                                                                                                                                                                                                 scattered radiation emitted by sample cell.
                     Type               Si-PIN semiconductor detector           420                                                           Additional Function Kit for EDX-LE
                     LN 2  Supply       Not required
                                                                                                                                              P/N 212-24714-42                                   3561 Universal X-Cell
                     Counting Method    Digital filter counting                                                                                Adds a general-analysis function to the EDX-LE. For details, please contact your
                                                                                                 1800                                         Shimadzu representative.                           P/N 219-85000-53 (100 pcs/set)
                                                                                                                                                                                                 (Outer diameter 31.6 mm, volume 8 mL)
                    Sample Chamber                                                                                                                                                               For liquid and thin-film samples. Equipped with a relief hole and
                     Measurement Atmosphere  Air                                                                                                                                                 liquid retainer in case of liquid expansion. Equipped with a ring for
                                                                                                                                                                                                 tightly holding thin-film samples with film.
                     Sample Observation  CCD camera
                    Data Processing Unit                                                                                                                                                         Polypropylene Film
                                                                                                                                                                                                 P/N 219-82019-05 (73 mm W × 92 m roll)
                     Memory             1 GB min.
                                                                                                                                                                                                 Sample-holding film. (For light element analysis)
                     HDD                80 GB min.
                     Resolution         1024 × 768 pixels min.
                     Printer            Color inkjet printer                                                                                                                                     Mylar Film
                                                                                                                                                                                                 P/N 202-86501-56 (500 sheets/set)
                     CD                 CD-ROM drive
                                                                                                                                                                                                 Sample-holding film. (For heavy element analysis)
                     OS                 Windows 7*
                    * Microsoft Office is not included in this OS.
                    Software
                     Screening Analysis  Simple operation software
                     Qualitative Analysis  Measurement/analysis software
                     Quantitative Analysis  Calibration curve method
                                        FP method
                                        Thin-film FP method (Option)
                                        BG-FP method (Option)
                     Matching Software  Option
                     Utilities          Automatic calibration functions (energy
                                        calibration, full-width half-maximum
                                        calibration)
                     Other Functions    System-status Monitoring Function
                                        Analysis-results Tabulation Function
                                        Analysis-results Report Creation Function

                     Installation Requirements

                                Guaranteed Performance  Guaranteed Operation
                     Temperature  10°C to 30°C (fluctuations   5°C to 35°C
                              should be 2°C/hour max.)
                     Humidity  40% to 70% (No condensation)  40% to 70% (No condensation)
                     Power Source  AC 100 V to 240 V ±10% 50/60 Hz, 150 VA grounded outlet
                              Power for peripheral devices (printer, PC, display monitor, etc.) must be
                              provided separately.











                                                                                                                                                                                                                                       EDX-LE
             14                                                                                                                                                                                                            Energy Dispersive X-ray Fluorescence Spectrometer  15
   9   10   11   12   13   14   15   16