Instrument Specifications
Primary Specifications Installation Example Options
Measurement Principle X-ray fluorescence spectrometry Dimensions of the Main Unit W 520 mm × D 650 mm × H 420 mm Halogen Screening Analysis Kit Sample Cells
Measurement Method Energy dispersive Main Unit Weight Approx. 60 kg P/N 212-24908-91 3571 General Open-End X-Cell (no lid)
Measurement Sample Type Solids, liquids, or powder This kit includes an instruction manual for Halogen analysis and a check sample required P/N 219-85000-55 (100 pcs/set)
for measurement of 6 elements (Cd, Pb, Hg, Cr, Br, and Cl) specified by the RoHS
Elements to be Detected 13 Al to 92 U directive and Halogen regulation. (Outer diameter: 31.6 mm, volume 10 mL)
Sample Chamber Size Max. W 370 mm × D 320 mm × H 155 mm Unit: mm Polyethylene sample cell used for liquid and powder samples. Used
with Mylar or polypropylene films.
At least 200 mm between unit and wall
X-Ray Generator RoHS, Halogen, and Antimony Screening Analysis Kit 3529 General X-Cell (with lid)
P/N 212-24922-91
X-Ray Tube Rh target
This kit includes an instruction manual and a check sample required for measurement of P/N 219-85000-52 (100 pcs/set)
Tube Voltage 5 kV to 50 kV 7 elements including those specified by the RoHS directive, Halogen regulation, and (Outer diameter: 32 mm, volume 8 mL)
Tube Current 1 A to 1,000 A 650 Antimony (Cd, Pb, Hg, Cr, Br, Cl, and Sb) Used for liquid samples. Equipped with relief hole and liquid retainer
Cooling Method Air cooling (with fan) in case of liquid expansion.
Exposure Area Automatic switching between 3, 5, and 10 mm Small Spot Solder Analysis Kit
dia. areas (1 mmø is an option) P/N 212-24850-41 3577 Micro X-Cell
Primary Filter Automatic switching between: 5 types + OPEN 520 (460) This kit includes an instruction manual for small spot solder analysis and a small spot P/N 219-85000-54 (100 pcs/set)
collimator plate required for measurement of a print circuit board. (Outer diameter 31.6 mm, volume 0.5 mL)
Detector For trace samples. Use with a collimator is recommended to reduce
scattered radiation emitted by sample cell.
Type Si-PIN semiconductor detector 420 Additional Function Kit for EDX-LE
LN 2 Supply Not required
P/N 212-24714-42 3561 Universal X-Cell
Counting Method Digital filter counting Adds a general-analysis function to the EDX-LE. For details, please contact your
1800 Shimadzu representative. P/N 219-85000-53 (100 pcs/set)
(Outer diameter 31.6 mm, volume 8 mL)
Sample Chamber For liquid and thin-film samples. Equipped with a relief hole and
Measurement Atmosphere Air liquid retainer in case of liquid expansion. Equipped with a ring for
tightly holding thin-film samples with film.
Sample Observation CCD camera
Data Processing Unit Polypropylene Film
P/N 219-82019-05 (73 mm W × 92 m roll)
Memory 1 GB min.
Sample-holding film. (For light element analysis)
HDD 80 GB min.
Resolution 1024 × 768 pixels min.
Printer Color inkjet printer Mylar Film
P/N 202-86501-56 (500 sheets/set)
CD CD-ROM drive
Sample-holding film. (For heavy element analysis)
OS Windows 7*
* Microsoft Office is not included in this OS.
Software
Screening Analysis Simple operation software
Qualitative Analysis Measurement/analysis software
Quantitative Analysis Calibration curve method
FP method
Thin-film FP method (Option)
BG-FP method (Option)
Matching Software Option
Utilities Automatic calibration functions (energy
calibration, full-width half-maximum
calibration)
Other Functions System-status Monitoring Function
Analysis-results Tabulation Function
Analysis-results Report Creation Function
Installation Requirements
Guaranteed Performance Guaranteed Operation
Temperature 10°C to 30°C (fluctuations 5°C to 35°C
should be 2°C/hour max.)
Humidity 40% to 70% (No condensation) 40% to 70% (No condensation)
Power Source AC 100 V to 240 V ±10% 50/60 Hz, 150 VA grounded outlet
Power for peripheral devices (printer, PC, display monitor, etc.) must be
provided separately.
EDX-LE
14 Energy Dispersive X-ray Fluorescence Spectrometer 15