Page 10 - EDX-7200
P. 10

Comprehensive Quantitation Functions






            Calibration Curve Method

            A standard sample is measured and the relationship with the fluorescent X-ray intensity
            plotted as a calibration curve, which is used for the quantitation of unknown samples.
            Although this method requires selection of a standard sample close to the unknown sample
            and creation of a calibration curve for each element, it achieves a high level of analysis
            accuracy.
            This method supports all types of corrections for coexistent elements,
            including absorption/excitation correction and correction for
            overlapping elements.

            Fundamental Parameter (FP) Method

            This method uses theoretical intensity calculations to determine the   Automatic Balance Setting Function (Patent pending)
            composition from the measured intensities. It's a powerful tool for   A balance setting is required to use the FP method on principal
            the quantitative analysis of unknown samples in cases where   components such as C, H, and O. The software automatically sets the
            preparation of a standard sample is difficult. (JP No. 03921872, DE   balance if it determines from the profile shape that a balance setting
            No. 60042990. 3-08, GB No. 1054254, US No. 6314158)  is required.

            Film FP Method
            The instrument also offers the thin-film FP method function. The film
            FP method permits the film thickness measurement of multilayer films,
            simultaneous film thickness measurements, and quantitative film
            composition.
            When using the film FP method, the substrate material, deposition
            sequence, and element information can be set.

            Background FP Method
                                                                                   Fluorescent X-ray  Compton-scattered
                                                                                   intensity of quantitative  radiation from Rh X-ray
            The background FP method adds scattered X-ray (background)             element        tube target material
            calculations to the conventional FP method, which only calculates the                 Rayleigh-scattered
            fluorescent X-ray peak intensity (net peak intensity).               Quantitative element  radiation from Rh
                                                                                                  X-ray tube target material
                                                                                 background
             (Patent pending : Japanese Patent No. 5975181)           X-ray intensity
            This method is effective at improving quantitation accuracy for small
                                                                                       Continuous X-rays
            quantities of organic samples, film thickness measurements of
            irregular-shaped plated samples, and film thickness measurements of
            organic films.                                                                Energy


            Matching Function
            The matching function compares analysis data for a sample with an
            existing data library and displays the results in descending degree of
            confidence.
            The library contains content data and intensity data and the user can
            register each type. The content data values can be entered manually.







                                                                                             Matching Results
                                                                                                                                                                                                                             EDX-7200
      10                                                                                                                                                                                                            Energy Dispersive X-Ray Fluorescence Spectrometer
   5   6   7   8   9   10   11   12   13   14   15