Page 19 - Shimadzu AA-6880
P. 19

$$        Installation Conditions



 Dimensions

 Wall       Make sure all required equipment is available before starting instrument installation. See the installation guidelines for details.
 Table
 150      200  Example of recommended piping for the atomic absorption system  Operating  Temperature range 10 to 35 °C
                                                                environment  Humidity range  20 to 80% (less than 70% when temperature is higher than 30 °C)
                          Stainless steel piping                                    1500 mm min. (W) × 700 mm min. (D)
                          (inner diameter greater than 7 mm)    Testing bench
                                                                                    Withstand load: 200 kg  *1, 2
 465                   Stop cock  Stop cock  Setting pressure 0.11 MPa  Material    Stainless  *3
 588                                                            Gas tube  Dimensions  7 mm min. I.D. × 80 mm min. (L)
 870                                                                    Acetylene   Cylinder: 0.11 MPa
              Pressure regulator                                                    Lab: 0.09 MPa
                                                                                    Compressor: 0.4 MPa
                                                                Gas supply  Air
                                            Acetylene gas
            Pressure range 0.3 MPa                              pressure            Lab: 0.35 MPa
            Setting pressure 0.09 MPa
                                       Setting pressure 0.4 MPa  (set secondary     Cylinder: 0.4 MPa
 280                                                            pressure)  Dinitrogen oxide  Lab: 0.35 MPa
 200 max.                                                                           Cylinder: 0.4 MPa
                                                                        Argon
                                                                                    Lab: 0.35 MPa
                                                                        Material    Stainless
                                                Setting pressure 0.4 MPa
                                         Dinitrogen                     Dimensions  For flame: Approx. 500 mm W × 500 mm D
             Pressure range 1 MPa
             Pressure range 1 MPa        oxide gas              Exhaust duct        For furnace: ø150 mm to ø200 mm
             Setting pressure 0.35 MPa
             Setting pressure 0.35 MPa
                                                                                    For flame: 600 to 1200 m 3 /hr
                                                                        Intake capacity
 ø150 to ø200                                         Argon gas                     For furnace: 10 to 180 m 3 /hr
             Pressure range 1 MPa                               Cooling water (GFA-6880)  Cooling water circulation unit or tap-water equipment  *4
             Setting pressure 0.35 MPa
                                                               *1 Atomic absorption spectrophotometer stand recommended
                                                Setting pressure 0.4 MPa
                                       Mist separator          *2 Maintain a free maintenance space of 150 to 200 mm to the sides and rear of the instrument.
 Exhaust fan                                                   *3 Do not use pipes containing copper, silver, gold, or mercury (or alloys containing these metals)
             Pressure range 1 MPa                                    as pipes for acetylene.
             Setting pressure 0.35 MPa                         *4 If tap-water equipment is used, ensure that it meets the specifications below.
                        Lab         Outdoor
                                                                Compatible faucet  Faucet 13 or 12 to 15 mm O.D. rimmed faucet
                            80 mm max.           Air compressor  Water temperature  10 to 30 °C
 Damper                                                         Water flow rate  0.6 to 1.5 L/min
                                                                Supply pressure  0.08 to 0.15 MPa  *5
                                                                Supply port  7 mm min.
                                                               *5 If the supply pressure exceeds 0.17 MPa, use the optional Regulator ASSY.
 Approx. 500 × 500
 417
 1100 to 1200

 Unit: mm
 AA-6880  S=1:20
 Above dimensions do not include PC or printer.
 702  GFA-6880
 ASC-6880
 390




 340
 702
 960
 Both for flame analysis and furnace analysis