Page 72 - Shimadu Consumables Catalog
P. 72
UHPLC/HPLC Columns
Shim-pack FC-ODS
High-speed analysis of 2,4-DNPH derivatized aldehydes / ketones
Conditions
Column : Shim-pack FC-ODS (75 mmL. × 4.6 mmI.D., 3 μm)
(P/N: 228-40511-92)
Mobile Phase : Water/Acetonitrile = 55/45 (v/v)
Flow Rate : 1.0 mL/min
Col. Temp. : 40 °C
Detection : UV 360 nm
Chromatogram of Standard Sample Chromatograms of Environmental Air (A)
(formaldehyde 0.35μg/mL, acetaldehyde and Indoor Air at Laboratory (B)
0.55μg/mL, 10μL injected)
Conditions
Column : Shim-pack FC-ODS
(75 mmL. × 4.6 mmI.D., 3 μm)
(P/N: 228-40511-92)
Mobile Phase : A) Water/Tetrahydrofuran = 8/2 (v/v)
B) Acetonitrile
A/B = 80/20 - 14min - 40/60 - 0.01min -
80/20 - 6min
Flow Rate : 1.2 mL/min
Col. Temp. : 40 °C
Detection : UV 365 nm
Chromatogram of Standard Sample Chromatograms of Indoor Air at Laboratory (A)
(each 0.3 ug/mL as carbonyl compounds, 10uL and Operation blank (B)
injection)
Product Information
Particle Size I.D. (mm) 2.0 4.6
(μm) Length (mm)
30 - 228-40511-91
3 75 228-40511-94 228-40511-92
150 228-40511-95 228-40511-93
Pg A-68