Page 7 - LCMS-TQ RX Series
P. 7

LCMS-TQ RX Series CoreSpray Technology



 Reliable  Resilient  Responsible



 CoreSpray

 Enhanced Ionization Performance


















 CoreSpray, a newly developed gas delivery system, improves
 performance and uniformity of nebulization with higher flows
 and better heat transfer.











 Peak Area Stability in Matrix: 20,000 Continuous Injections  PERFORMANCE CONCIERGE ™
 CoreSpray technology applied to the repeated analysis of pesticides in black tea matrix resulting in enhanced   PERFORMANCE CONCIERGE makes tuning a mass
 robustness (%RSD less than 3.5%, 20,000 injections). Data generated without internal standards and without   spectrometer easier than ever before. A tuning
 diverting to waste for the initial polar matrix effects.  standard is automatically introduced into the
               instrument to verify parameters, including mass
               accuracy, resolution, and signal strength. Based on
               these checks, tuning may be automatically initiated
 Q 192.10>160.15 (+)  Q 192.10>160.15 (+)  Q 192.10>160.15 (+)
               to ensure optimal performance. Should the criteria
 1.5e5  1.5e5  1.5e5
 1.3e5  1.3e5  1.3e5
               not be met, PERFORMANCE CONCIERGE will
 1.0e5  1.0e5  1.0e5
               diagnose the issue and alert the operator of required
 7.5e4  7.5e4  7.5e4
               maintenance and maximize uptime.
 5.0e4  5.0e4  5.0e4
 2.5e4  2.5e4  2.5e4
 0.0e0  0.0e0  0.0e0
 3.9  4.0  4.1  4.2  4.3  4.4  3.9  4.0  4.1  4.2  4.3  4.4  3.9  4.0  4.1  4.2  4.3  4.4
 Injection 1  Injection 10,000  Injection 20,000
 8.E+05  %RSD
 Carbendazim  3.26%
 6.E+05                                      Analytical Intelligence represents Shimadzu’s innovative approach to automated analytical
                                             instrument optimization. This concept encompasses a suite of systems and software
 Peak area  4.E+05  Spinetoram L  2.92%      designed to emulate the decision-making process of expert operators. It autonomously
                                             assesses the instrument status, helps to give actionable feedback and where it can it fixes
 2.E+05                                      problems. By bridging the gap between varying levels of user expertise and instrument
 Acetamiprid  2.52%                          familiarity, Analytical Intelligence significantly enhances the reliability of your data.
 0.E+00
 0  5,000  10,000  15,000  20,000
 Injection number


 6                                                                                                                 7
   2   3   4   5   6   7   8   9   10   11   12