                                                                                  It is recommended to use a solvent delay column (installed after

                                                                                  the mixer and before the autosampler) to delay the elution of
                                                                                  PFAS originating from solvent bottles and other parts of the liquid

                                                                                  chromatography system (e.g., pumps and tubing). As shown in
                                                                                  Figure 3 below, using the delay column enables the detection of

                                                                                  PFOA originating solely from the sample.



                                                                                  Additionally, bypassing the degasser when possible is recommended
                                                                                  as well as replacing any PTFE-containing tubing and parts in the LC.







                                                                                                   Figure 3: Chromatogram of PFOA:

                                                                                              (a) without delay column and (b) with delay column























                                                                                                                                                                 10