Page 15 - Shimadzu PDA-8000
P. 15
Specifications
Spectrometer Unit Optical system Paschen–Runge mount
Monochromator With temperature control
Excitation Unit Excitation unit REM power supply
Emission table Argon gas atmosphere
Software No. of processing channels 64 channels max.
Standard sample registration 10000 samples max.
Matrix element correction 32 elements max./element
Correction 100 % correction
Master curve correction
Analysis function Content analysis (four-time analysis, round analysis)
Analysis processing Recycling
Additional analysis
Canceling
Ending
Display/print order and number of digit settings
Analysis result transmission RS-232C
LAN (TCP/IP)
Analysis results display X ¯ -R control chart
Histogram
Maintenance Maintenance period control
Password Protection of analysis information
Installation and External Dimensions and
Optional Accessories Weight
Installation Environment PDA Instrument : W1690 × D1066 × H1142 mm
Temperature : 15 °C to 30 °C PC Set : W600 × D528 × H1380 mm
Humidity : 70 % max. Weight : Approx. 600 kg
Power Supply : Single phase 200 V ±10 % 50/60 Hz 1.5 kVA * For installation details, refer to the Pre-Installation
Requirement.
Installation : Separate, 30 max.
Argon Gas : Purity 99.999 % min. Dew point −70 °C max. Wall
A switching system with 2 or more 7m cylinders is
3
600 min.
convenient
Optional Accessories
(Installation spaces and power supplies are required separately.)
• Sampling Mold and Surface Plate
• L-1000 Bench lathe : Single phase 100 V 300 W
• Belt Grinder with Dust Collector : 3-phase 200 V 1150 W 1066
• MT-11M Counter Electrode Grinder : Single phase 100 V 100 W
1690 600 min.
600 PC set
units in mm
528
15