Page 4 - Shimadzu ON-LINE TOC-VCSH
P. 4

C391-E081A


               Specifications                                                                                          ON-LINE TOC-VCSH

               Measured items          NPOC (TOC by IC subtraction using acid sparging), TC, IC, TOC (TC-IC), TN (optional TNM-1 required)
                                       680°C combustion catalytic oxidation/NDIR method
               Measurement method
                                       720°C combustion catalytic oxidation/ chemiluminescence method (when TNM-1 is connected)
                                       TC: 0-500 µg/L to 0-25,000 mg/L variable
               Measurement range
                                       IC: 0-500 µg/L to 0-30,000 mg/L variable
               Measurement cycle       Approx. 5 to 999 minutes (for NPOC measurement)
               Number of measurement channels  1 channel                                                                                 On-Line TOC Analyzer TOTAL ORGANIC CARBON ANALYZER
               Reproducibility         CV 1.5% max.
               Linearity               ±2% F.S. max.                                                                                     ON-LINE TOC-VCSH
               Zero stability          ±2% F.S. max. per day
               Span stability          ±2% F.S. max. per day (ambient temperature fluctuations within 5°C)
               Calibration cycle       Variable in the range 1 to 999 hours
               Sample injection method  Automatic injection using syringe pump/slider
               Sample injection volume  Variable in the range 10 to 2,000 µL
               Sample dilution function  Diluted in syringe; dilution factor: 2 to 50
               IC pretreatment         Automatic acidification and sparging
                                       High-purity air or oxygen (1 ppm max. of CO, CO2, or HC)
                                       High-purity nitrogen (1 ppm max. of CO, CO2, or HC) (when using optional nitrogen gas carrier kit)
               Carrier gas             Supply pressure: Approx. 600 kPa
                                       Flow rate: 150 mL/min (230 mL/min with sparging)
                                       Analog output (1 output): Select from 4 to 20 mA, 0 to 16 mA (insulation, load resistance: 750 Ωmax.), and 0 to 1VDC
               Measurement value output
                                       RS-232C: Baud rate; 9,600 bps, Data length; 8 bits, Parity; none, Stop bits; 1
               Alarm output            Measurement value upper limit and upper upper limit alarms,
                                       system error alarm: No-voltage contact output (maximum: 5VA, 0.1A, 50V)
               Control input           Remote stopping and starting of sample measurement and calibration
                                       Flow rate: Approx. 0.1 to 1 L/min
               Sample conditions       Sample volume: 5 to 7 mL per measurement
                                       Temperature: 0 to 90°C
               Ambient temperature     5 to 35°C
                                       AC 100 - 127V ±10%    MAX    800VA     50/60 Hz
               Power supply            AC 200 - 240V ±10%    MAX    1200VA   50/60 Hz
               Dimensions              Approx. W440 x D560 x H460 mm (excluding protrusions)
               Weight                  Approx. 40 kg
               Special Accessories

                                    Part Number                             Description
               OCT-1                638-93150-01  Permits the automated measurement of up to seven samples, in addition to an on-line flow line.
               OCT-1 (second unit)  638-93150-11  Second unit to expand OCT-1. Permits the automated measurement of up to 15 samples,
                                                 in addition to an on-line flow line.
               TNM-1                638-91065    Permits TN (total nitrogen) measurements.
               Nitrogen carrier gas kit  638-42054  High-purity N2  gas (1 ppm max. of CO, CO2, or HC) can be used as the carrier gas.
                                                 *Simultaneous use of TNM-1 and this option is not possible.
                                                 *The measuring range with this option becomes 0-500 µg/L to 0-100 mg /L for TC and IC both.
               Sample adjustment tank  638-42053  Used when bubbles are contained within the sample.

               External Dimensions Diagram

                    ON-LINE TOC-VCSH + Sample adjustment tank




                                            460


                  Sample
             adjustment tank
                                 440
                               (depth 560)




                                               Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its
                                               affiliates, whether or not they are used with trademark symbol “TM” or “®”.
                                               Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services. Shimadzu
                                               disclaims any proprietary interest in trademarks and trade names other than its own.
                                               For Research Use Only. Not for use in diagnostic procedures.
                                               The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice. Shimadzu
                                               does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
       www.shimadzu.com/an/                                                                        © Shimadzu Corporation, 2012
                                                                                                Printed in Japan 3655-06230-10AIT