C391-E081A


 Specifications  ON-LINE TOC-VCSH

 Measured items  NPOC (TOC by IC subtraction using acid sparging), TC, IC, TOC (TC-IC), TN (optional TNM-1 required)
 680°C combustion catalytic oxidation/NDIR method
 Measurement method
 720°C combustion catalytic oxidation/ chemiluminescence method (when TNM-1 is connected)
 TC: 0-500 µg/L to 0-25,000 mg/L variable
 Measurement range
 IC: 0-500 µg/L to 0-30,000 mg/L variable
 Measurement cycle  Approx. 5 to 999 minutes (for NPOC measurement)
 Number of measurement channels  1 channel  On-Line TOC Analyzer TOTAL ORGANIC CARBON ANALYZER
 Reproducibility  CV 1.5% max.
 Linearity  ±2% F.S. max.  ON-LINE TOC-VCSH
 Zero stability  ±2% F.S. max. per day
 Span stability  ±2% F.S. max. per day (ambient temperature fluctuations within 5°C)
 Calibration cycle  Variable in the range 1 to 999 hours
 Sample injection method  Automatic injection using syringe pump/slider
 Sample injection volume  Variable in the range 10 to 2,000 µL
 Sample dilution function  Diluted in syringe; dilution factor: 2 to 50
 IC pretreatment  Automatic acidification and sparging
 High-purity air or oxygen (1 ppm max. of CO, CO2, or HC)
 High-purity nitrogen (1 ppm max. of CO, CO2, or HC) (when using optional nitrogen gas carrier kit)
 Carrier gas  Supply pressure: Approx. 600 kPa
 Flow rate: 150 mL/min (230 mL/min with sparging)
 Analog output (1 output): Select from 4 to 20 mA, 0 to 16 mA (insulation, load resistance: 750 Ωmax.), and 0 to 1VDC
 Measurement value output
 RS-232C: Baud rate; 9,600 bps, Data length; 8 bits, Parity; none, Stop bits; 1
 Alarm output  Measurement value upper limit and upper upper limit alarms,
 system error alarm: No-voltage contact output (maximum: 5VA, 0.1A, 50V)
 Control input  Remote stopping and starting of sample measurement and calibration
    Flow rate: Approx. 0.1 to 1 L/min
 Sample conditions  Sample volume: 5 to 7 mL per measurement
    Temperature: 0 to 90°C
 Ambient temperature  5 to 35°C
 AC 100 - 127V ±10%    MAX    800VA     50/60 Hz
 Power supply  AC 200 - 240V ±10%    MAX    1200VA   50/60 Hz
 Dimensions  Approx. W440 x D560 x H460 mm (excluding protrusions)
 Weight  Approx. 40 kg
 Special Accessories

 Part Number  Description
 OCT-1  638-93150-01  Permits the automated measurement of up to seven samples, in addition to an on-line flow line.
 OCT-1 (second unit)  638-93150-11  Second unit to expand OCT-1. Permits the automated measurement of up to 15 samples,
 in addition to an on-line flow line.
 TNM-1  638-91065  Permits TN (total nitrogen) measurements.
 Nitrogen carrier gas kit  638-42054  High-purity N2  gas (1 ppm max. of CO, CO2, or HC) can be used as the carrier gas.
 *Simultaneous use of TNM-1 and this option is not possible.
 *The measuring range with this option becomes 0-500 µg/L to 0-100 mg /L for TC and IC both.
 Sample adjustment tank  638-42053  Used when bubbles are contained within the sample.

 External Dimensions Diagram

 ON-LINE TOC-VCSH + Sample adjustment tank




 460


 Sample
 adjustment tank
 440
 (depth 560)




 Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its
 affiliates, whether or not they are used with trademark symbol “TM” or “®”.
 Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services. Shimadzu
 disclaims any proprietary interest in trademarks and trade names other than its own.
 For Research Use Only. Not for use in diagnostic procedures.
 The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice. Shimadzu
 does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
 www.shimadzu.com/an/  © Shimadzu Corporation, 2012