Page 3 - Shimadzu iSpect DIA-10
P. 3

Comprehensive Particle Characterization  Reliable Particle Detection System




 Shimadzu’s iSpect DIA-10 Dynamic Image Analyzer combines particle size and image analysis technology   The system utilizes a microcell and advanced optics to accurately and ef ciently detect particles. Using normal
 to offer complete particle characterization. It can perform particle imaging, size analysis, and foreign   lenses, the apparent size of particles can be affected by the depth of the particle relative to the lens. The iSpect
 object detection, and obtain size distributions and number concentration, in as little as two minutes.  DIA-10 uses a telecentric lens that maintains a constant image magni cation. This means that no matter where the
               particle is located in the  eld of vision, the system will accurately determine the size of the particle. The autofocus
               function increases the imaging ef ciency  which makes it possible to accurately detect foreign objects and obtain
                                                 *1
               repeatable number concentration.
 Particle images  *1: The ratio of the number of sample particle images acquired to the number of particles in the sample that has  owed through the system.

                Ef cient Image Acquisition

               The microcell increases image acquisition ef ciency by passing particles through a narrow imaging area, which
               optimizes the number of particles observed. Compared with a conventional cell and lens, the microcell clearly
 Particle shape  Number   shows particle images and fewer particles pass outside of the lateral imaging area. This makes it possible to
 iSpect  ™  concentration  *2: When measured using a Shimadzu standard sample. Depends on the sample.
               reliably detect particles and obtain highly repeatable number concentration (Coef cient of Variation (CV) ±5 % ).
                                                                                                            *2

               ■ Microcell                                       ■ Conventional Cell

                                                                       Particles passing
                                      Pass through the imaging area    outside the      Pass outside the imaging area
 Particle size   Foreign object       ⇒Most particles detected         imaging area     ⇒ Blurred and missed particles
 distribution  detection
                                          Cell width 1 mm                                   Cell width
                                                                                            10 mm min.


 Features  ● Reliable particle detection system  Cell depth                                   Cell depth

 ● Unprecedented functionality and performance  0.12 mm                                       1 mm max.

 ● Easy operation
                     Sharp, in-focus
                        image         Red frame: Imaging area            Blurred        Red frame: Imaging area
                                                                        particles


               ■ Example of Number Concentration Measurement
                                    3500

                                    3000

                                   Measured Value (particles/mL)  2000
                                    2500


                                    1500

                                    1000

                                     500

                                      0
                                       0      500   1000   1500  2000   2500   3000   3500
                                                        Theoretical Value (particles/mL)
                                Results of 6 Measurements (The ● mark is the mean, and the error bar is the standard deviation)
   1   2   3   4   5   6   7   8