P. 2
Comprehensive Particle Characterization Reliable Particle Detection System
Shimadzu’s iSpect DIA-10 Dynamic Image Analyzer combines particle size and image analysis technology The system utilizes a microcell and advanced optics to accurately and ef ciently detect particles. Using normal
to offer complete particle characterization. It can perform particle imaging, size analysis, and foreign lenses, the apparent size of particles can be affected by the depth of the particle relative to the lens. The iSpect
object detection, and obtain size distributions and number concentration, in as little as two minutes. DIA-10 uses a telecentric lens that maintains a constant image magni cation. This means that no matter where the
particle is located in the eld of vision, the system will accurately determine the size of the particle. The autofocus
function increases the imaging ef ciency which makes it possible to accurately detect foreign objects and obtain
*1
repeatable number concentration.
Particle images *1: The ratio of the number of sample particle images acquired to the number of particles in the sample that has owed through the system.
Ef cient Image Acquisition
The microcell increases image acquisition ef ciency by passing particles through a narrow imaging area, which
optimizes the number of particles observed. Compared with a conventional cell and lens, the microcell clearly
Particle shape Number shows particle images and fewer particles pass outside of the lateral imaging area. This makes it possible to
iSpect ™ concentration *2: When measured using a Shimadzu standard sample. Depends on the sample.
reliably detect particles and obtain highly repeatable number concentration (Coef cient of Variation (CV) ±5 % ).
*2
■ Microcell ■ Conventional Cell
Particles passing
Pass through the imaging area outside the Pass outside the imaging area
Particle size Foreign object ⇒Most particles detected imaging area ⇒ Blurred and missed particles
distribution detection
Cell width 1 mm Cell width
10 mm min.
Features ● Reliable particle detection system Cell depth Cell depth
● Unprecedented functionality and performance 0.12 mm 1 mm max.
● Easy operation
Sharp, in-focus
image Red frame: Imaging area Blurred Red frame: Imaging area
particles
■ Example of Number Concentration Measurement
3500
3000
Measured Value (particles/mL) 2000
2500
1500
1000
500
0
0 500 1000 1500 2000 2500 3000 3500
Theoretical Value (particles/mL)
Results of 6 Measurements (The ● mark is the mean, and the error bar is the standard deviation)