Page 4 - Shimadzu IRXross
P. 4
High-End Sensitivity for Countless Applications
The IRXross is a mid-level FTIR model that achieves high-end level S/N. It enables best-in-class low noise with P-P
values of 55,000:1 for one minute of integration.
Moving mirror
Beam splitter
Fixed mirror
Astoundingly Low Noise
The 100 %T line was obtained by successively measuring the background and sample values without placing a
sample in the sample compartment. Excluding the peaks for water vapor and carbon dioxide, the noise level (P-P
value) was within ±0.005 %T, which shows that it can acquire data with low noise.
105.0
%T
100.010
± 0.005%T 102.5
%T
100.005
100.000 100.0
99.995
97.5
99.990
2200 2190 2180 2170 2160 2150 2140 2130 2120 2110 2100
cm -1
95.0
4000 3500 3000 2500 2000 1750 1500 1250 1000 cm -1
IRXross 100 %T Line