Page 9 - Shimadzu IG-1000 Plus
P. 9

Detailed Visualization of Reliable Data






 Measurement Data  Analysis of Mixed Samples


 The following data was obtained by analyzing various sample particle groups of various sizes.  The IG method uses the diffusion of the diffraction grating created by particles; therefore, the
            signal size does not depend on the particle size. This means that evaluation of mixed samples is
            possible. With methods based on scattered light, even if the volume is the same, the signal size is

 Fullerene Hydroxide  Particle Size Distribution (Volumetric Basis)  proportional to the cube of the particle diameter, making evaluation of mixed samples difficult.
 Normalized Particle Amount
 Fullerene hydroxide, a typical material of samples in the single
 nano region, can be analyzed with high reproducibility.  Colloidal silica  Particle Size Distribution (Volumetric Basis)




                                                  Normalized Particle Amount

 Particle Diameter (nm)
 (Sample provided by Prof. Kokubo of Osaka University)






                                                                       Particle Diameter (nm)
 Silica Sample with Broad Distribution  Particle Size Distribution (Volumetric Basis)             : A  : B  : A+B
 Even with samples covering a wide distribution, there is no bias
 Normalized Particle Amount
 toward larger particles, and the existence of smaller particles is   Polystyrene latex  Particle Size Distribution (Volumetric Basis)
 accurately captured.







 Particle Diameter (nm)                           Normalized Particle Amount





                                                                        Particle Diameter (nm)
 Samples Containing Contamination  Particle Size Distribution (Volumetric Basis)     : 50 nm  : 100 nm  : 50 nm+100 nm
 Measurement results are not affected by minute quantities of
 Normalized Particle Amount
 contamination. (The example shows the results obtained for a
 sample with a distribution centering on a diameter of 50 nm
 that contains a 1% concentration of 1-µm particles.)






 Particle Diameter (nm)






                                                                                              IG-1000 Plus
 8                                                                                              Single Nano Particle Size Analyzer  9
   4   5   6   7   8   9   10   11   12   13   14