Page 4 - Shimadzu ICPS-7510 - New
P. 4

Applicable to all types of analytical assessments






                        Resolution                                                                                                    Stable analysis with
                                                                                                                                      a vacuum type spectrometer
                        0.0066nm
                         0.013nm                                                                                                      The inside of the spectrometer is a vacuum. This
                                                                                                                                      enables highly sensitive analysis of Al, P, S, B, I,
            High Resolution                                                                                                           Br, Hg etc. which have intense spectra in the
                                                                                                                                      vacuum ultraviolet region.
            The resolution is high, at 0.0066nm (160 to 458nm) and            Cu 213.598                                              A gas purge is unnecessary, which means that
            0.013nm (458 to 850nm).                                                                                                   there is no fluctuation due to convection of the
            The effect of adjacent spectra is minimal, which enables                                                                  purge gas, and no distortion of the optical
            high precision trace analysis.                                                                                            system. This leads to a short stabilization time,
                                                                                                                                      and the continuance of stable analysis over a long
                                                                                P 213.620
                                                                                                                                      period of time.
                                                                                              The ICPS-7510                           The condenser lens can be cleaned in the
                                                                                              sharply resolves                        vacuum, which means that the internal parts of   Limits of detection(3   )
                                                                                              the Cu 213.598nm and                    the spectrometer will not be exposed to the   Element  Wavelength    g/L
                                                                                              the P 213.620nm.
                                                                                                                                      atmosphere.                                     AI      167.079    0.3
                                                                                                                                                                                      P       177.499    2.0
                                                                                                                                                                                      S       180.731    4.0
                                                                                                                                                                                      * Using the standard nebulizer (measurement in the radial direction)


                Holographic double grating
             For short wavelengths: 3600groves / mm
             For long wavelengths: 1800groves / mm
                     160nm    850nm                                                                                                   Highly sensitive analysis
                                                                                                                                      through axial observation
            Coversa wide range of
            spectral measurements                                                                                                     With the attachment for observing the light in
                                                                                                                                      the axial direction that comes as standard, the
            A holographic double grating with 3600 grooves/mm                                                                         ICPS-7510 is configured for ultra trace analysis.
            for short wavelengths and 1800 grooves/mm for long                                                                        Combined with the observation of the plasma in
            wavelengths is installed. A wide range of spectra from                                                                    the radial direction, analyses can be made from
            160nm to 850nm can be measured at will under                                                                              ultra trace amounts to high concentrations.
            computer control.
            High precision analysis is possible for elements that
            have intense spectra in the vacuum ultraviolet region -
            Al, P, S, B – as well as for elements that have their
            spectra in the near infrared region such as Li, K, and Rb.
                                                                                                                                                                                            Limits of detection (3   )
                                                                                                                                                                                            Element  Wavelength    g/L
                                                                                                                                                                                              Pb      220.351     2.0
                                                                                                                                                                                              Cd      226.502     0.3
                                                                                                                                                                                              A s     189.042     4.0
                                                                                                                                                                                             * Using the standard nebulizer










                                                                                                                                                                                                                             ICPS-7510
     6                                                                                                                                                                                                                  Sequential Plasma Emission Spectrometer  7
   1   2   3   4   5   6   7   8   9