Page 4 - Shimadzu ICPS-7510 - New
P. 4
Applicable to all types of analytical assessments
Resolution Stable analysis with
a vacuum type spectrometer
0.0066nm
0.013nm The inside of the spectrometer is a vacuum. This
enables highly sensitive analysis of Al, P, S, B, I,
High Resolution Br, Hg etc. which have intense spectra in the
vacuum ultraviolet region.
The resolution is high, at 0.0066nm (160 to 458nm) and Cu 213.598 A gas purge is unnecessary, which means that
0.013nm (458 to 850nm). there is no fluctuation due to convection of the
The effect of adjacent spectra is minimal, which enables purge gas, and no distortion of the optical
high precision trace analysis. system. This leads to a short stabilization time,
and the continuance of stable analysis over a long
P 213.620
period of time.
The ICPS-7510 The condenser lens can be cleaned in the
sharply resolves vacuum, which means that the internal parts of Limits of detection(3 )
the Cu 213.598nm and the spectrometer will not be exposed to the Element Wavelength g/L
the P 213.620nm.
atmosphere. AI 167.079 0.3
P 177.499 2.0
S 180.731 4.0
* Using the standard nebulizer (measurement in the radial direction)
Holographic double grating
For short wavelengths: 3600groves / mm
For long wavelengths: 1800groves / mm
160nm 850nm Highly sensitive analysis
through axial observation
Coversa wide range of
spectral measurements With the attachment for observing the light in
the axial direction that comes as standard, the
A holographic double grating with 3600 grooves/mm ICPS-7510 is configured for ultra trace analysis.
for short wavelengths and 1800 grooves/mm for long Combined with the observation of the plasma in
wavelengths is installed. A wide range of spectra from the radial direction, analyses can be made from
160nm to 850nm can be measured at will under ultra trace amounts to high concentrations.
computer control.
High precision analysis is possible for elements that
have intense spectra in the vacuum ultraviolet region -
Al, P, S, B – as well as for elements that have their
spectra in the near infrared region such as Li, K, and Rb.
Limits of detection (3 )
Element Wavelength g/L
Pb 220.351 2.0
Cd 226.502 0.3
A s 189.042 4.0
* Using the standard nebulizer
ICPS-7510
6 Sequential Plasma Emission Spectrometer 7