Page 14 - Shimadzu HMV-G Series
P. 14

global w430×h280



                  Specifications



                                                        HMV-G31 Series                 HMV-G30 Series                                                                               HMV-G31-FA and HMV-G31-XY (HMV-G31 + Electric XY Stage)
                                Monochrome Camera   HMV-G31S     HMV-G31D        HMV-G30S       HMV-G30D                                                  Monochrome Camera
                                without Electric Turret  344-04273-˘˘  344-04274-˘˘  344-04271-˘˘  344-04272-˘˘                                           without Electric Turret
                                Monochrome Camera   HMV-G31ST    HMV-G31DT      HMV-G30ST      HMV-G30DT                                                  Monochrome Camera   HMV-G31-FA-S  HMV-G31-FA-D  HMV-G31-XY-S  HMV-G31-XY-D
                                with Electric Turret  344-04277-˘˘  344-04278-˘˘  344-04275-˘˘  344-04276-˘˘                                              with Electric Turret  344-04279-˘˘  344-04280-˘˘  344-04281-˘˘  344-04282-˘˘
                    Model P/N                                                                                                                Model P/N
                                Color Camera without   HMV-G31S-HC  HMV-G31D-HC                                                                           Color Camera without
                                Electric Turret  344-04287-˘˘   344-04288-˘˘                                                                              Electric Turret
                                Color Camera with   HMV-G31ST-HC  HMV-G31DT-HC                                                                            Color Camera with   HMV-G31-FA-S-HC  HMV-G31-FA-D-HC  HMV-G31-XY-S-HC  HMV-G31-XY-D-HC
                                Electric Turret  344-04289-˘˘   344-04290-˘˘                                                                              Electric Turret  344-04291-˘˘  344-04292-˘˘    344-04293-˘˘   344-04294-˘˘
                                                                       16 stages in total                                                                                                       16 stages in total
                                                   9.807 mN, 19.61 49.03, 98.07, 245.2, 490.3, 980.7 mN, 1.96 N, 2.942, 4.903, 9.807, 19.61 N                                9.807 mN, 19.61 49.03, 98.07, 245.2, 490.3, 980.7 mN, 1.96 N, 2.942, 4.903, 9.807, 19.61 N
                    Test Force                      12 stages (HV0.001, HV0.002, HV0.005, HV0.01, 0.025, 0.05, 0.1, 0.2, 0.3, 0.5, 1, and 2)   Test Force                    12 stages (HV0.001, HV0.002, HV0.005, HV0.01, 0.025, 0.05, 0.1, 0.2, 0.3, 0.5, 1, and 2)
                                                                  plus four user-designated stages  *1                                                                                      plus four user-designated stages  *1
                    Loading Unit                              Automatic loading and automatic force changing                                 Loading Unit                              Automatic loading and automatic force changing
                    Test Force Duration Time              User-designated settings from 0 to 999 sec *2 , in 1 sec increments                Test Force Duration Time              User-designated settings from 0 to 999 sec *2 , in 1 sec increments
                    Maximum Number of Indenter Attachments  1    2× (1 empty)      1           2× (1 empty)                                  Maximum Number of Indenter Attachments  1       2              1               2
                    Standard Indenter Provided                          7JDLFST *OEFOUFS                                                     Standard Indenter Provided                          Vickers Indenter
                    Maximum Number of Objective Lens Attachments  2  4             2              4                                          Maximum Number of Objective Lens Attachments  2  4             2               4
                    Standard Objective Lenses Provided  40× (1 empty)  40× 10× (2 empty)  40× (1 empty)  40× 10× (2 empty)                                                40× (Super-long   40× (Super-long
                                              Automatic readings via digital image analysis, and manual                                      Standard Objective Lenses Provided  Working Distanse)  Working Distanse)  40× (1 empty)  40× 10× (2 empty)
                    Indentation Reading System                                 Manual readings via an optical microscope                                                    5× (0 empty)   5× (2 empty)
                                              settings for line length measurement in the PC window
                                                                                                                                                                                        Automatic readings via digital image analysis,
                    Eyepiece                                                              ×10                                                Indentation Reading System
                                                                                                                                                                                  and manual settings for line length measurement in the PC window
                                                     USB Connection (USB 2.0)                                                                Eyepiece
                                                       1/3-inch image size
                                 Monochrome
                                                  Measurement image 640 × 480 pixels                                                                                                          USB Connection (USB 2.0)
                    Digital Camera               256 tint gradations (digital image data)                                                                 Monochrome                            1/3-inch image size
                                                                                                                                                                                           Measurement image 640 × 480 pixels
                                                  USB Connection (USB 3.0 Super Speed)                                                                                                    256 tint gradations (digital image data)
                                 Color                 2/3-inch image size                                                                   Digital Camera
                                                 Measurement image 2448 × 2048 pixels                                                                                                     USB Connection (USB 3.0 Super Speed)
                                                                                                                                                          Color                                 2/3-inch image size
                    Effective Measurement Range    Monochrome Camera: 120 × 90 µm                                                                                                         Measurement image 2448 × 2048 pixels
                    (For 40× objective lens)         Color Camera: 195 × 160 µm         250×250µm
                                                                                                                                             Effective Measurement Range                   Monochrome Camera: 120 × 90 µm
                                                     Area: 100 × 100 mm; Stroke: ±12.5 mm; Sample: Maximum height 100 mm                     (For 40× objective lens)                        Color Camera: 195 × 160 µm
                                                     Instrument’s lateral direction (width direction): Unlimited *3 ;
                                                     Instrument’s depth direction: Unlimited for samples with a width of 120 mm max. *3                                Area: 125 × 125 mm; XY Axis Stroke: ±25 mm; Sample: Maximum height 65 mm
                    XY Stage
                                                     When sample width is greater than 120 mm: 200 mm max. *3                                                          Instrument’s lateral direction (width direction): Unlimited
                                                     Z Axis Stroke: 60 mm; Spacer provided: Thickness 40 mm; Maximum Loading: 10 kg          XY Stage                  Instruments depth direction: Unlimited for samples with a width of 120 mm max. *3
                                                     The center of gravity of objects loaded must be located on the stage.                                             When sample width is greater than 120 mm: 175 mm max.
                                                                                                                                                                       Z Axis Stroke: 45 mm; Spacer provided: Thickness 30 mm
                    Autofocus Function                                                                                                                                 Note: There is also a large sample stage edition with an area of 200 × 200 mm and an XY axis stroke of ±60 mm.
                    Pattern Settings Items
                                                                                                                                             Autofocus Function                      Y                              N
                    Image Acquisition Functions                                                                                                                          Linear division, linear pitch, matrix, rectangular division, rectangular pitch, square, circular division,
                                              ᶃ Vickers Hardness HV ᶄ Knoop hardness HK *4  ᶅ Brinell hardness: HB *4  ᶆ Triangular Pyramid Hardness HT *3  Pattern Settings Items  circular pitch, arbitrary, zigzag
                                              ᶇ Length, direct reading L (µm) *5  ᶈ Fracture Toughness Kc
                    Data Processing Functions                                                                                                Image Acquisition Functions            Total image acquisition function and edge detection function
                                              ᶉ Shape correction function (cylindrical surfaces and spherical surfaces)
                                              Note: The cracks for measuring fracture toughness are determined manually by the operator.                                ᶃ Vickers Hardness HV ᶄ Knoop hardness HK *4  ᶅ Brinell hardness: HB *4  ᶆ Triangular Pyramid Hardness HT *3
                                                   Data No., diagonal line length, hardness, conversion value, average, standard deviation,   Data Processing Functions  ᶇ Fracture Toughness Kc ᶈ Shape correction function (cylindrical surfaces and spherical surfaces)
                                 Displayed Items                                                                                                                        Note: The cracks for measuring fracture toughness are determined manually by the operator.
                    Results Displayed              coefficient of variation, maximum value, minimum value, pass/fail determination
                                 Graphical Display  Variance graph, depth of hardening graph *6  Variance graph, histogram, depth of hardening graph *6   Displayed Items   Data No., diagonal line length, hardness, conversion value, average, standard deviation,
                                 USB                                        Transfer of test results to a USB flash drive (CSV format)       Results Displayed              coefficient of variation, maximum value, minimum value, pass/fail determination
                    External Output                                                    5IFSNBM QSJOUFS                                                    Graphical Display              Variance graph, depth of hardening graph *6
                                 Printer           Printer output from a PC via software
                                                                                   $PNQBUJCMF XJUI MBTFS QSJOUFST                            External Output  USB
                    External Dimensions                           Approx. W350 × D570 × H540 mm                                                           Printer                          Printer output from a PC via software
                    Weight                                              "QQSPY     LH                                                        External Dimensions                            Approx. W350 × D570 × H540 mm
                                                                                                                                             Weight                                              Approx. 50 kg
                                              Main unit: Single-phase 100 to 115 VAC, 230 VAC, 100 VA (50/60 Hz)
                    Power Supply              Note: Prepare a grounded outlet. (D type ground, with a ground resistance of 100 Њ max.)                                       Main unit: Single-phase 100 to 115 VAC, 230 VAC, 100 VA (50/60 Hz)
                                              Note: With the HMV-G31 series, a power supply for the PC will be required in addition to the above mentioned.  Power Supply    Motor Control Box: Single-phase 100 VAC, 100 VA (50/60 Hz)
                                                                                                                                                                             Note: Prepare a grounded outlet. (D type ground, with a ground resistance of 100 Њ max.)
                                               OS: Windows ®  10 (64 bit edition)                                                                                            Note: A power supply for the PC will be required in addition to the above mentioned.
                                                    ®
                                               CPU: Intel  Core ™  i5-6500 or faster
                                                    *8
                                               USB Ports : For monochrome camera                                                                                                 OS: Windows  10 (64 bit edition)
                                                                                                                                                                                         ®
                    PC Provided *8                    Uses two USB 2.0 or 3.0 ports                                                                                              CPU: Intel ®  Core ™  i5-6500 or faster
                                                      For color camera                                                                                                           USB Ports: For monochrome camera
                                                      Uses one USB 3.0 port (required) and                                                   PC Provided *8                            Uses two USB 2.0 or 3.0 ports
                                                      one USB 2.0 or 3.0 port
                                                                                                                                                                                       For color camera
                  *1) An arbitrary test force can be configured in the range of 9.807 mN to 19.61 N, in increments of the minimum 9.807 mN (HV 0.001).                                 Uses one USB 3.0 port (required) and one USB 2.0 or 3.0 port
                  *2) When performing tests with the retention time set to 60 seconds or longer, turn ON the instrument at least 30 minutes before starting the test.
                  *3) The width of open space in the instrument’s depth direction is approximately 140 mm.                                 *5) The direct length reading is only available with HMV-G30 series.
                    When testing samples with a width of 120 mm or less and a depth of 175 mm or more, give consideration to the required movement range of the XY stage,   *6) Depth of hardening graphs can only be displayed when the digital micrometer (optionally available) is used.
                    and ensure that the width of the open space is not exceeded.                                                           *7) The set items with the PC provided by Shimadzu is recommended.
                    Ensure that the shape of the sample is such that it is stable when placed on the XY stage.                             *8) Another USB 2.0 or 3.0 port must be added if using the electric micrometer kit.
                  *4) The Knoop indenter, Brinell indenter and triangular pyramidal indenter are optional. The Brinell hardness and triangular pyramidal hardness are only read   Connect the electric micrometer kit to a USB 2.0 port when using it with a color camera type unit.
                    in manually.
                                                                                                                                                                                                                                HMV-G
                                                                                                                                                                                                                             Micro Vickers Hardness Tester
      14                                                                                                                                                                                                                                      15
   9   10   11   12   13   14   15   16