Page 8 - Shimadzu eTOC series
P. 8
Reliability to support regulations
Measurement reliability supports regulatory compliance requirements Security functionality for supporting data management
The UV oxidation-conductivity method enables highly sensitive measurements with a simple analyzer design.
The system achieves high sensitivity with a detection limit of 0.1 µg/L, which is suitable for ultra-pure water quality control. User authentication and access rights control
It is also compliant with the pharmacopoeia requirements of various countries, such as the USP and EP.
Users can be authenticated based on an ID and password.
User access rights are controlled according to four user levels: “Administrator”, “Main
User”, “User”, or “Guest”.
Internal structure and measurement principle
Operation/Event log
Switch OFF the lamp, start the sample
Sample
To drain IN ow, and measure the conductivity of All operations performed by operators, measurement value alarms, instrument errors,
the unoxidized sample. and other operating events are recorded.
Filters by "Category", such as "Measurement Alarms" or "Instrument Errors", can be
To applied for browsing the log history.
detector Stop the sample ow and switch the Descriptions and reasons for frequently performed operations can be registered in
lamp ON to oxidize organic matter in the
Conductivity advance and then recalled later.
sensor sample.
Pump Excimer Integrated data management
lamp Switch OFF the lamp and start the sample
Thermo ow to measure the conductivity of the A daily report summarizing measurement results from each day (p.7) and an operation log can be exported daily to a LabSolutions
-meter
oxidized sample. database via a network.
Data from multiple eTOC analyzers and data from other analytical instruments can be managed centrally together.
Calculate the TOC value based on the
conductivity difference between oxidized
and unoxidized samples.
Drain Drain
eTOC
Measurement results Operation
(daily report) log
Stable and Accurate Data Measurement
8.0
LC
7.0
6.0
5.0
ʢµg/Lʣ 4.0 eTOC
TOC
3.0
2.0
LabSolutions
database
1.0
0.0
0:09:00 0:39:00 1:09:00 1:39:00 2:09:00 2:39:00 3:09:00 3:39:00 4:09:00 4:39:00 5:09:00 5:39:00 6:09:00 6:39:00 7:09:00 7:39:00 8:09:00 8:39:00 9:09:00 9:39:00 10:09:00 10:39:00 11:09:00 11:39:00 12:09:00 12:39:00 13:09:00 13:39:00 14:09:00 14:39:00 15:09:00 15:39:00 16:09:00 16:39:00 17:09:00 17:39:00 18:09:00 18:39:00 19:09:00 19:39:00 20:09:00 20:39:00 21:09:00 21:39:00 22:09:00 22:39:00 23:09:00 23:39:00
AA
Time
24-Hour Continuous eTOC Measurement Data of Ultra-Pure Water
It shows that results are stable enough to check the quality of ultra-pure water around 5 g/L TOC. TOC-L UV
eTOC series
8 On-Line TOC Analyzer for Pure Water 9