Page 11 - Shimadzu eTOC series
P. 11
Options Routine Consumable Parts Speci cations*
Description P/N Remarks Description P/N Remarks Model TOC-1000e
Vial Sampler 638-57230-41 Four 40-mL vials Excimer Lamp 638-69201-41 Measurement Items TOC, conductivity (or speci c resistance), and temperature
Enables mounting to Pump Head 638-59384-41 Measurement Principle UV oxidation-conductivity method
Bracket Kit 638-23033-41
either a wall or pole. Sample Filter Element 040-22305-80 Temp.: 10 to 50 °C
Enables analog output Printer Paper 078-15046-51 3 pcs Sample Water Conditions Flowrate: 30 to 500 mL/min
External I/O Kit 638-79217-41
and contact input/output. Conductivity: 2 S/cm or lower
Power Supply Measurement Cycle 2.5 min, 5 min, 10 min, 15 min, 30 min, 1 hr, 2 hrs, 4 hrs, 12 hrs, or 24 hrs
638-68186-41
Terminal Block Kit TOC: 0 to 2000 µg/L
60-µm element for 1/8-inch Measurement Range Conductivity: 0.023 to 206 µS/cm (without temperature correction)
Sample Filter 040-22305-45
OD tubing Temp.: 10 to 50 °C
Needle Valve for For adjusting sample owrate Detection Limit TOC: 0.1 µg/L
Adjusting Sample Flowrate 040-22308-03 in 1/8-inch OD tubing TOC: ±5 % (500 µg/L sucrose)
Used during maintenance, Accuracy Conductivity: ±2 % (147 µS/cm at 25 °C)
Sample Flow Shutoff Valve 040-07042-01
etc. with 1/8-inch OD tubing Temp.: ±0.5 °C (at 25 °C)
Tubing OD adapter from Repeatability TOC: CV ≤ 1 % or SD ≤ 0.3 g/L, whichever is greater
Reducer Union 035-60414-50
1/8 to 1/4-inch Linearity TOC: R ≥ 0.98
2
Calibration* 2 Creates a 4-point calibration curve automatically
3
Outputs USB ash drive, printer, LAN ( le or Internet), Modbus (TCP/IP), analog* , or contact* 3
Conditioned Standard Solutions
Analog Output* 3 4 to 20 mA output, three channels from TOC, conductivity, speci c resistance, or temperature
Application P/N Remarks Alarm: Measurement value alarm, device alarm, serious failure, or power supply interruption
Contact Output* 3
Blank Water 638-60252-91 Four vials of blank water for thermometer validation and rinsing Event: Ready for measurement, measuring, and analog synchronization signal
Set of 147 µS/cm Potassium Chloride Contact Input* 3 Start measurement
638-60254-91 Two vials of 147 µS/cm KCl and one vial of blank water for conductivity calibration
Standard Conductivity Solutions Display 7-inch color touch panel
Set of 0-250-500 µg/L Sucrose Standard One vial of blank water and one vial each of 250 and 500 µg/L sucrose for TOC Ingress Protection Rating IP33
638-60255-91
TOC Solutions calibration and validation Indoors, with 10 to 40 °C ambient temperatures
Set of 0-500 µg/L Sucrose Standard TOC Installation Site Requirements Tabletop, wall-mounted* or pole-mounted* 4
4
638-60256-91 One vial of blank water and one vial of 500 µg/L sucrose for TOC calibration
Solutions Power Supply 100 to 240 V AC, 50/60 Hz
Set of 0-500-1000-2000 µg/L Sucrose 638-60257-91 One vial of blank water and one vial each of 500, 1000, and 2000 µg/L sucrose for TOC Dimensions W270 × D140 × H180 mm
Standard TOC Solutions calibration and validation Weight Approx. 2.88 kg
Set of 0-2000 µg/L Sucrose Standard TOC
638-60258-91 One vial of blank water and one vial of 2000 µg/L sucrose for TOC calibration *1: The speci cation above is available when managed and operated appropriately by trained people.
Solutions *2: Using an optional vial sampler
USP 643 System Suitability Test Set, One vial of blank water, one vial of 500 µg/L sucrose, and one vial of 500 µg/L *3: Using an optional external I/O kit
638-60259-91 *4: Using an optional bracket kit
Con rmed (USP Standard) 1,4-benzoquinone for TOC system suitability test
USP 643 System Suitability Test Set, 638-60260-91 One vial of blank water, one vial of 500 µg/L sucrose, and one vial of 500 µg/L
Equivalent 1,4-benzoquinone for TOC system suitability test
Installation Space
TOC-1000e TOC-1000e with Vial Sampler (Optional)
Wall-mounted Pole-mounted Bracket Wall-mounted Pole-mounted
eTOC series
10 On-Line TOC Analyzer for Pure Water 11