Page 11 - Shimadzu eTOC series
P. 11

Options  Routine Consumable Parts  Speci cations*
 Description  P/N  Remarks  Description  P/N  Remarks  Model  TOC-1000e
 Vial Sampler  638-57230-41 Four 40-mL vials  Excimer Lamp  638-69201-41  Measurement Items  TOC, conductivity (or speci c resistance), and temperature
 Enables mounting to  Pump Head  638-59384-41  Measurement Principle  UV oxidation-conductivity method
 Bracket Kit  638-23033-41
 either a wall or pole.  Sample Filter Element  040-22305-80  Temp.: 10 to 50 °C
 Enables analog output  Printer Paper  078-15046-51  3 pcs  Sample Water Conditions  Flowrate: 30 to 500 mL/min
 External I/O Kit  638-79217-41
 and contact input/output.                  Conductivity: 2  S/cm or lower
 Power Supply   Measurement Cycle           2.5 min, 5 min, 10 min, 15 min, 30 min, 1 hr, 2 hrs, 4 hrs, 12 hrs, or 24 hrs
 638-68186-41
 Terminal Block Kit                         TOC: 0 to 2000 µg/L
 60-µm element for 1/8-inch  Measurement Range  Conductivity: 0.023 to 206 µS/cm (without temperature correction)
 Sample Filter  040-22305-45
 OD tubing                                  Temp.: 10 to 50 °C
 Needle Valve for  For adjusting sample  owrate  Detection Limit  TOC: 0.1 µg/L
 Adjusting Sample Flowrate  040-22308-03  in 1/8-inch OD tubing  TOC: ±5 % (500 µg/L sucrose)
 Used during maintenance,  Accuracy         Conductivity: ±2 % (147 µS/cm at 25 °C)
 Sample Flow Shutoff Valve 040-07042-01
 etc. with 1/8-inch OD tubing               Temp.: ±0.5 °C (at 25 °C)
 Tubing OD adapter from  Repeatability       TOC: CV ≤ 1 % or SD ≤ 0.3  g/L, whichever is greater
 Reducer Union  035-60414-50
 1/8 to 1/4-inch  Linearity                 TOC: R  ≥ 0.98
                                                 2
                Calibration* 2              Creates a 4-point calibration curve automatically
                                                                                           3
                Outputs                     USB  ash drive, printer, LAN ( le or Internet), Modbus (TCP/IP), analog* , or contact* 3
 Conditioned Standard Solutions
                Analog Output* 3            4 to 20 mA output, three channels from TOC, conductivity, speci c resistance, or temperature
 Application  P/N  Remarks                  Alarm: Measurement value alarm, device alarm, serious failure, or power supply interruption
                Contact Output* 3
 Blank Water  638-60252-91  Four vials of blank water for thermometer validation and rinsing  Event: Ready for measurement, measuring, and analog synchronization signal
 Set of 147 µS/cm Potassium Chloride   Contact Input* 3  Start measurement
 638-60254-91  Two vials of 147 µS/cm KCl and one vial of blank water for conductivity calibration
 Standard Conductivity Solutions  Display   7-inch color touch panel
 Set of 0-250-500 µg/L Sucrose Standard   One vial of blank water and one vial each of 250 and 500 µg/L sucrose for TOC   Ingress Protection Rating  IP33
 638-60255-91
 TOC Solutions  calibration and validation  Indoors, with 10 to 40 °C ambient temperatures
 Set of 0-500 µg/L Sucrose Standard TOC   Installation Site Requirements  Tabletop, wall-mounted*  or pole-mounted* 4
                                                             4
 638-60256-91  One vial of blank water and one vial of 500 µg/L sucrose for TOC calibration
 Solutions      Power Supply                100 to 240 V AC, 50/60 Hz
 Set of 0-500-1000-2000 µg/L Sucrose   638-60257-91  One vial of blank water and one vial each of 500, 1000, and 2000 µg/L sucrose for TOC   Dimensions  W270 × D140 × H180 mm
 Standard TOC Solutions  calibration and validation  Weight  Approx. 2.88 kg
 Set of 0-2000 µg/L Sucrose Standard TOC
 638-60258-91  One vial of blank water and one vial of 2000 µg/L sucrose for TOC calibration  *1: The speci cation above is available when managed and operated appropriately by trained people.
 Solutions  *2: Using an optional vial sampler
 USP 643 System Suitability Test Set,   One vial of blank water, one vial of 500 µg/L sucrose, and one vial of 500 µg/L   *3: Using an optional external I/O kit
 638-60259-91  *4: Using an optional bracket kit
 Con rmed (USP Standard)  1,4-benzoquinone for TOC system suitability test
 USP 643 System Suitability Test Set,   638-60260-91  One vial of blank water, one vial of 500 µg/L sucrose, and one vial of 500 µg/L
 Equivalent  1,4-benzoquinone for TOC system suitability test
 Installation Space
 TOC-1000e  TOC-1000e with Vial Sampler (Optional)


















 Wall-mounted  Pole-mounted  Bracket  Wall-mounted                   Pole-mounted

                                                                                                eTOC series
 10                                                                                             On-Line TOC Analyzer for Pure Water  11
   6   7   8   9   10   11   12