Page 7 - Shimadzu DSC-60 Plus Series
P. 7
Sample pans
ᶅ P/N
ᶃ201-52943 Al ʢAluminium) crimp pans + lids, ø6 × 1.5 (50/set)
ᶄ201-51976 Platinum pan, ø6 × 2.5
ᶅ201-56927 Platinum lid, ø6
ᶃ ᶄ ᶆ ᶇ ᶈ ᶉ ᶆ201-54321 Alumina pan, ø6 × 2.5
ᶇ201-53102-84 Nickel pans, ø6 × 2 (50/set)
ᶈ201-58294-90 Copper pans, ø6 × 1.5 (50/set)
ᶉ201-54439 Quartz pan, ø6 × 2.5
ᶊ201-53090 Al hermetic pans, ø6 × 1.6 (50/set), limit pressure: 0.3 MPa
ᶋ222-02067-92 Pressure-proof stainless steel hermetic pans ø6 × 5 (50/set), limit pressure: 5 MPa Max.500°C
ᶊ ᶋ ᶌ ᶌ222-13073-91 Pressure-proof Al hermetic pans ø6 × 5 (50/set), limit pressure: 5 MPa Max.300°C
ᶍ201-57268-90 Al macro pans, ø6 × 5 (50/set)
Other pans
ᶍ
ᶎ346-66963-91 Al crimp pans and lids for autosampler, ø6 × 3 (100/set)
ᶏ346-68518-91 Al hermetic pans for autosampler (DSC) ,
limit pressure: 0.3MPa, ø4.4 × 4 (100/set)
ᶐ346-68796-91 Al hermetic pans for autosampler (DTG) ,
limit pressure: 0.3MPa, ø6 × 4 (100/set)
ᶑ346-68334-91 Copper pans for Autosampler, ø6 × 3 (100/set) Max.300°C
DSC-60 Plus Specifications
Measurement principle Heat-flux type
1
Temperature range -140 to 600°C * (With standard cooling chamber and using liquid nitrogen )
Calorimetric measurement range ±150mW
Noise level Less than 0.5μW (rms, when held at 150°Cʣ
Sample quantity (pan quantity) Standard AL crimp pan approx.40μl
Programmable heating/cooling rate +/- 0.1 to +/- 99.9°C /min ( 0.1°C /hr or 0.1°C /min step)
Programmable hold time 1min - 999hr (1min or 1hr step)
Atmosphere Nitrogen, inert gas, dry air gas flow
Interface RS-232C, 1
Control software Thermal analysis workstation TA-60WS
Power supply AC100V, 120V, 230V 50/60Hz MAX 800VA * 2
Size W:320 x H:500 x D:500 (mm)
Weight 28kg
*1 Temperature range of Automatic Cooling model (TAC/L) is -130-600°C. The measurement temperature is different when using an automatic cooling unit.
*2 This indicates only the power supply for the main unit. Installation of the stabilized power supply is recommended according to the power supply circumstances.
Pre Installation Requirements
Analytical balances Gas
To weigh the sample, prepare an analytical balance which Purge gas (atmospheric gas to be used)
allows precise reading up to 0.01 mg. Cleaning air tank or air compressor
Ìi®ÊUÊ/Ê«iÀvÀÊV}Êi>ÃÕÀiiÌÊÜÌ
ÊÌ
iÊ
DSC, dry gas (nitrogen or dry air) is additionally
required.
Other UÊ*Ài«>ÀiÊ>ÊÌ>]Ê«ÀiÃÃÕÀiÊÀi`ÕViÀÊ>`Ê}>ÃÊvÜÊÀ>ÌiÊ
regulator separately.
Do not install the device in a place exposed to
direct sunlight, a place exposed to direct wind from an air
conditioner, a dusty place, a place subject to large
vibrations, or a place subject to large temperature
fluctuation.
DSC-60 Plus Series
Differential Scanning Calorimeter 7