Page 7 - Shimadzu DSC-60 Plus Series
P. 7

Sample pans







                      ᶅ                            P/N
                                               ᶃ201-52943  Al ʢAluminium) crimp pans + lids, ø6 × 1.5 (50/set)
                                               ᶄ201-51976  Platinum pan, ø6 × 2.5
                                               ᶅ201-56927  Platinum lid, ø6
                  ᶃ   ᶄ   ᶆ   ᶇ   ᶈ   ᶉ        ᶆ201-54321  Alumina pan, ø6 × 2.5
                                               ᶇ201-53102-84  Nickel pans, ø6 × 2 (50/set)
                                               ᶈ201-58294-90  Copper pans, ø6 × 1.5 (50/set)
                                               ᶉ201-54439  Quartz pan, ø6 × 2.5
                                               ᶊ201-53090  Al hermetic pans, ø6 × 1.6 (50/set), limit pressure: 0.3 MPa
                                               ᶋ222-02067-92  Pressure-proof stainless steel hermetic pans ø6 × 5 (50/set), limit pressure: 5 MPa Max.500°C
                  ᶊ      ᶋ      ᶌ              ᶌ222-13073-91  Pressure-proof Al hermetic pans ø6 × 5 (50/set), limit pressure: 5 MPa Max.300°C
                                               ᶍ201-57268-90  Al macro pans, ø6 × 5 (50/set)
                                               Other pans
                                        ᶍ
                                               ᶎ346-66963-91  Al crimp pans and lids for autosampler, ø6 × 3 (100/set)
                                               ᶏ346-68518-91  Al hermetic pans for autosampler (DSC) ,
                                                          limit pressure: 0.3MPa, ø4.4 × 4 (100/set)
                                               ᶐ346-68796-91  Al hermetic pans for autosampler (DTG) ,
                                                          limit pressure: 0.3MPa, ø6 × 4 (100/set)
                                               ᶑ346-68334-91  Copper pans for Autosampler, ø6 × 3 (100/set) Max.300°C

            DSC-60 Plus Specifications

               Measurement principle       Heat-flux type
                                                     1
               Temperature range            -140 to 600°C *  (With standard cooling chamber and using liquid nitrogen )
               Calorimetric measurement range  ±150mW
               Noise level                 Less than 0.5μW (rms, when held at 150°Cʣ
               Sample quantity (pan quantity)  Standard AL crimp pan approx.40μl
               Programmable heating/cooling rate   +/- 0.1 to +/- 99.9°C /min ( 0.1°C /hr or 0.1°C /min step)
               Programmable hold time      1min - 999hr (1min or 1hr step)
               Atmosphere                  Nitrogen, inert gas, dry air gas flow
               Interface                   RS-232C, 1
               Control software            Thermal analysis workstation TA-60WS
               Power supply                AC100V, 120V, 230V 50/60Hz MAX 800VA * 2
               Size                        W:320 x H:500 x D:500 (mm)
               Weight                      28kg
            *1 Temperature range of Automatic Cooling model (TAC/L) is -130-600°C. The measurement temperature is different when using an automatic cooling unit.
            *2 This indicates only the power supply for the main unit. Installation of the stabilized power supply is recommended according to the power supply circumstances.
            Pre Installation Requirements



              Analytical balances                                Gas

              To weigh the sample, prepare an analytical balance which   Purge gas (atmospheric gas to be used)
              allows precise reading up to 0.01 mg.              Cleaning air tank or air compressor
                                                                  œÌi®ÊUÊ/œÊ«iÀvœÀ“ÊVœœˆ˜}ʓi>ÃÕÀi“i˜ÌÊ܈̅Ê̅iÊ
                                                                       DSC, dry gas (nitrogen or dry air) is additionally
                                                                       required.
              Other                                                   UÊ*Ài«>ÀiÊ>ÊÌ>˜Ž]Ê«ÀiÃÃÕÀiÊÀi`ÕViÀÊ>˜`Ê}>ÃÊvœÜÊÀ>ÌiÊ
                                                                       regulator separately.
              Do not install the device in a place exposed to
              direct sunlight, a place exposed to direct wind from an air
              conditioner, a dusty place, a place subject to large
              vibrations, or a place subject to large temperature
              fluctuation.



                                                                                           DSC-60 Plus Series
                                                                                                 Differential Scanning Calorimeter  7
   2   3   4   5   6   7   8