Page 4 - AXIMA Performance
P. 4

AXIMA Performance™ -

            Excellence without Compromise


                                                         High performance MS data
             %Int.                     ~147300
             100
             90                                          The AXIMA Performance demonstrates high resolution
             80
                                                         and mass accuracy across a wide mass range, from
             70
                                                         pharmaceutical compounds, through peptides to high
             60
             50           73954                          mass proteins, to enable a variety of applications in the
             40                                          research environment. A highly optimized linear mode
             30                                          generates reproducible and sensitive results for very
             20                                          high mass compounds and complexes, extending the
             10                                          array of sample classes that may be analyzed. A typical
              0
                      50000    100000    150000   200000  example of an intact high mass protein,
                                 m/z
                                                         Immunoglobulin G, is shown (left).
            Linear MS spectrum of Immunoglobulin G
                                                         Powerful precursor ion selection

                                                         The revolutionary monoPULSE ion gate allows the
                                                         selection of MS/MS precursors with industry-leading
                                                         resolution. Ions from complex mixtures or closely
                                                         associated neighbouring isotopic envelopes may be
                                                         isolated and subsequently fragmented. The gate
                                                         resolution of greater than 400 (FWHM) readily permits
                                                         the analysis of peptides with similar nominal mass,
                                                         even with overlapping isotopic distributions.

                                                         Efficient MS/MS from
                                                         high-energy CID

                                                         Once selected for MS/MS fragmentation, precursor
                                                         ions are subjected to true high-energy collisions with
                                                         the chosen collision gas, e.g. helium. The resulting
                                                         fragment ions are analyzed in the second TOF region
                                                         incorporating the curved field reflectron and detected
                                                         with the ultrafast MCP detector.
                                                         The high sensitivity means that good quality, usable
                                                         MS/MS spectra can often be obtained for low sample
                                                         amounts.
              %Int.
              100
                                                         The novel additional Low Mass Zoom (LMZ) scan allows
              90
                                                         focus of the analysis to be directed around the low
              80
                                                         mass region of the MS/MS spectrum. This feature can
              70
                                                         be employed to increase confidence in MS/MS based
              60
                                                         quantitation experiments and to aid in de novo
              50
                                                         sequencing of unknowns by enhancing the immonium
              40
                                                         ion region of the spectrum.
              30
              20
              10
              0
                  100  200  300  400  500  600  700  800  900  1000
                                  m/z
            Typical high-energy CID MS/MS spectrum








                                                                       AXIMA Performance
   1   2   3   4   5   6