            No.   21




             Serotonin (5-hydroxytryptamine) is a biologically active substance that plays an important role in the body, functioning
             in the blood to constrict the vascular smooth muscle, and to promote platelet aggregation. This article introduces an
             example of ultrafast analysis of serotonin in the blood utilizing the Nexera UHPLC system and the RF-20Axs high-
             sensitivity fluorescence detector.
             Analysis of Serotonin in the Blood

             The blood sample was subjected to deproteinization via an aqueous trichloroacetic acid solution. A Shim-pack XR-
             ODS III (2 mm internal diameter, 50 mm length) with a 1.6 µm particle size was used, and detection was performed via
             the RF-20Axs fluorescence detector. The maximum system load pressure in this analysis was approximately 79 MPa.




                                                              Column     : Shim-pack XR-ODSⅢ
                                                                          (50 mmL. x 2.0 mmI.D., 1.6 µm)
                                                              Pressure   : 79 MPa
                                                              Mobile Phase   : 0.15 mmol/L Acetate buffer (pH 4.7) / Methanol
                                                                         = 9 / 1 (v / v)
                                                              Flow Rate   : 0.7 mL/min
                                                              Column Temp. : 37 ℃
                                                              Injec  n Volume : 1 µL
                                                              Detec  n   : Fluorescence (RF-20AXS) Ex. 300 nm, Em. 350 nm
                                                              Cell Temp.   : 25℃
                                                              Flow Cell   : Semi-micro cell


                                                              Peak :
                                                              1. Serotonin













                 * This data was provided by BML, Inc.
                 * The data in this document was not acquired by instruments approved by the Japanese Pharmaceutical Affairs Law.











                                                                                                     First Edition: June, 2012


                                             For Research Use Only. Not for use in diagnostic procedures.
                                             The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
                                             The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
                                             accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
                                             use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
                                             to change without notice.
                                                                                                 © Shimadzu Corporation, 2012
   4   5   6   7   8   9   10   11   12   13   14