Page 8 - Shimadzu-MassSpectrometerCatalog
P. 8

Mass spectrometers must first ionize sample molecules.


            Many type of ionization have been developed.
            Electron ionization (EI) ionizes samples using accelerated thermal electrons,
            70 eV. EI is generally used for samples, whose molecular weight is less than
                                                                  High Polarity
            1000. It is typically used in combination with a gas chromatograph to ana-       LCMS
            lyze highly volatile molecules or gases.
            Electrospray ionization (ESI) ionizes samples in solution by spraying them. It   Middle Polarity
            is mainly used in combination with liquid chromatographs.
                                                                           GCMSS                 MALDI
            Matrix Assisted Laser Desorption/Ionization (MALDI) desorbs sample mole-
                                                                   Low Polarity
            cules by laser energy and ionizes those by transferring charge from matrix
                                                                          100  1000   5000   10000   100000
            ions. MALDI can apply wide range of samples by changing matrix reagents,
                                                                                     Molecular Weight
            hence, this ionization method apply to synthetic polymer and complicated
            samples.





            ESI (Electrospray Ionization)                       ESI        From column


                                                                                    ESI probe
            A sample solution is introduced to capillary applying high electric
                                                                                              (1) Capillary
            voltage and sprayed at the edge of capillary. Droplets of sample
            solutions are charged and sample molecules are ionized in liquid      (2)         (2) Nebulizer gas
            phase.                                                       (1)                  (3) Dry gas
            ESI is best suited to ionizing medium-to-high polarity substances,
                                                                                              (4) DL (desolvation line)
            peptides, proteins, and oligonucleotides.
                                                                         (3)   (4)





            DUIS (ESI+APCI)                                     DUIS       From column

                                                                                    ESI probe
            The DUIS-2020 is a dual ion source that ionizes samples using both
                                                                                              (1) Nebulizer gas
            ESI and APCI (atmospheric pressure chemical ionization) modes.
            This is best suited to ionizing low-to-medium polarity substances.    (1)         (2) Corona needle
                                                                                              (3) Dry gas
                                                                                              (4) DL (desolvation line)
                                                                         (2) (3)   (4)






            MALDI                                                              : Sample molecules
                                                                               : Matrix
            (Matrix Assisted Laser                                             : Positive ions
                                                                               : Negative Ions
             Desorption/Ionization)
                                                                 Laser
            After applying the sample and matrices solution on a sample target     Vacuum
            and dry up, laser irradiates to sample spot. The sample and matrix
            molecules are desorbed and ionized, rapidly.
                                                                               Sample
            MALDI enables to ionize high molecular weight compounds, such
                                                                              Sample target           Sample target
            as proteins and polymers without dissociation.


      8
   3   4   5   6   7   8   9   10   11   12   13