Page 12 - Shimadzu XSeeker 8000
P. 12

C251-E042
               Main Specifications                                                                                      XSeeker 8000


                   X-Ray Generator Output                                   160 kV 1.2 mA
                   X-Ray Detector                                         Flat panel detector
                   Max. Input Resolution                                Approx. 5.6 million pixels
                                                     SRD*1 = 475 mm      ⼾300 mm×H 320 mm,10 kg (Including sample setting jig)
                   Max. Sample Size/Weight
                                                      SRD = 550 mm       П150 mm×H 320 mm,10 kg (Including sample setting jig)
                                                                                                                                         Benchtop X-Ray CT System
                                                      SRD = 475 mm                  П85 mm×H 70 mm
                   Max. CT-Scan Region
                                                      SRD = 550 mm                  П100 mm×H 80 mm                                      XSeeker 8000
                   Spatial Resolution                                   100  m (SRD = 550 mm)
                   CT Data Acquisition Time                                12 sec or 5 min
                              Main Unit                       Single phase 100 to 240 V AC ± 10 %, 50/60 Hz, 1.0 kVA
                   Rated
                   Output     Control Computer                Single phase 100 to 240 V AC ± 10 %, 50/60 Hz, 1.0 kVA
                              Grounding                       D type grounding (Grounding resistance of 100   max.)
                   Weight                                                    290 kg
                   Operating Environment Conditions  Ambient temperature: 10 to 30 °C; Ambient humidity: RH of 45 to 80 % max. (No condensation)
                   External Leakage Dose                                   1  Sv/h or less.
               *1 SRD: The source-to-rotation center distance (SRD) is the distance from the X-ray source to the rotation center of the sample.

               XSeeker 8000 Image and Layout (Units: mm)

                               (1950)
                  (350*)  Main Unit
                                                  (526)
                (1100)
                  (750)  (650)                 (728)
                                                  (700)                     (702)
                           (893)       (820)
                (150*)     (900)
                    CT System Table (Option)  Desk for Control PC (Option)
                 Recommended Installation Space: W1950 × D1100 mm
                 Note: Leave at least 350 mm of space behind the CT system table and 150mm of space left side in order to permit access for maintenance work.





                • Automated support functions utilizing digital technologies, such as M2M, IoT, and
                 Artificial Intelligence (AI), that enable higher productivity and maximum reliability.
                • Allows a system to monitor and diagnose itself, handle any issues during data acquisition
                 without user input, and automatically behave as if it were operated by an expert.
                • Supports the acquisition of high quality, reproducible data regardless of an operator's skill
                 level for both routine and demanding applications.



               XSeeker and the Analytical Intelligence logo are trademarks of Shimadzu Corporation or its af liated companies in Japan and/or other countries.
               VGSTUDIO MAX is a trademark of Volume Graphics GmbH.






                                                 For Research Use Only. Not for use in diagnostic procedures.
                                                 This publication may contain references to products that are not available in your country. Please contact us to check the availability of
                                                 these products in your country.
                                                 Company names, products/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation,
                                                 its subsidiaries or its af liates, whether or not they are used with trademark symbol “TM” or “®”.
                                                 Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether
                                                 or not they are used with trademark symbol “TM” or “®”.
                                                 Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
         www.shimadzu.com/an/                    The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice.
                                                 Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
                                                                    © Shimadzu Corporation, 2022 / First Edition: May 2022, 3655-03203-PDFIK, C251-E042
   7   8   9   10   11   12