Page 1 - Shimadzu XSeeker 8000
P. 1

C251-E042
 Main Specifications  XSeeker 8000


 X-Ray Generator Output   160 kV 1.2 mA
 X-Ray Detector  Flat panel detector
 Max. Input Resolution  Approx. 5.6 million pixels
 SRD*1 = 475 mm  ⼾300 mm×H 320 mm,10 kg (Including sample setting jig)
 Max. Sample Size/Weight
 SRD = 550 mm  П150 mm×H 320 mm,10 kg (Including sample setting jig)
               Benchtop X-Ray CT System
 SRD = 475 mm  П85 mm×H 70 mm
 Max. CT-Scan Region
 SRD = 550 mm  П100 mm×H 80 mm  XSeeker 8000
 Spatial Resolution  100  m (SRD = 550 mm)
 CT Data Acquisition Time  12 sec or 5 min
 Main Unit  Single phase 100 to 240 V AC ± 10 %, 50/60 Hz, 1.0 kVA
 Rated
 Output  Control Computer  Single phase 100 to 240 V AC ± 10 %, 50/60 Hz, 1.0 kVA
 Grounding  D type grounding (Grounding resistance of 100   max.)
 Weight  290 kg
 Operating Environment Conditions  Ambient temperature: 10 to 30 °C; Ambient humidity: RH of 45 to 80 % max. (No condensation)
 External Leakage Dose  1  Sv/h or less.
 *1 SRD: The source-to-rotation center distance (SRD) is the distance from the X-ray source to the rotation center of the sample.

 XSeeker 8000 Image and Layout (Units: mm)

 (1950)
 (350*)  Main Unit
 (526)
 (1100)
 (750)  (650)  (728)
 (700)  (702)
 (893)  (820)
 (150*)  (900)
 CT System Table (Option)  Desk for Control PC (Option)
 Recommended Installation Space: W1950 × D1100 mm
 Note: Leave at least 350 mm of space behind the CT system table and 150mm of space left side in order to permit access for maintenance work.





 • Automated support functions utilizing digital technologies, such as M2M, IoT, and
 Artificial Intelligence (AI), that enable higher productivity and maximum reliability.
 • Allows a system to monitor and diagnose itself, handle any issues during data acquisition
 without user input, and automatically behave as if it were operated by an expert.
 • Supports the acquisition of high quality, reproducible data regardless of an operator's skill
 level for both routine and demanding applications.



 XSeeker and the Analytical Intelligence logo are trademarks of Shimadzu Corporation or its af liated companies in Japan and/or other countries.
 VGSTUDIO MAX is a trademark of Volume Graphics GmbH.






 For Research Use Only. Not for use in diagnostic procedures.
 This publication may contain references to products that are not available in your country. Please contact us to check the availability of
 these products in your country.
 Company names, products/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation,
 its subsidiaries or its af liates, whether or not they are used with trademark symbol “TM” or “®”.
 Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether
 or not they are used with trademark symbol “TM” or “®”.
 Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
 www.shimadzu.com/an/  The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice.
 Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
 © Shimadzu Corporation, 2022 / First Edition: May 2022, 3655-03203-PDFIK, C251-E042
   1   2   3   4   5   6