Page 3 - Shimadzu TGA-50 Series
P. 3
Measurement can be performed in various gases
and high-capacity cells can be used.
In the gas flow line of the TGA-50 Series, active gas can be introduced
from ᶃ shown in the figure on the right while protecting the balance
unit by feeding inert gas or other material from ᶄ. A compact balance
ᶄ
mechanism and small heating furnace are employed for the minimum
dead volume and higher air tightness and gas displacement efficiency.
In addition, various types of measurements can be performed by using
high-capacity cells specially designed for the suspension method. ᶃ
TGA
mg
Mass loss: -0.26 mg
0.00
-1.00
TG gas flow line system
0.00 20.00 40.00 60.00
Time (min)
or
p
T
rans
he
t
W
Va
ater
miss
f
lm
or
oods
F
Rate
ion
of
i
aF
em
Meas
ur
of
Measurement of the Water Vapor Transmission Rate of a Film for Foods
ent
(using the TGA-51 and special cell)
TGA
%
100.1
100.0 -0.167%
99.9
99.8
99.7
50.00 100.00 150.00
Temp (%)
Measurement of the Moisture Content of PET Fiber (using a quartz crucible)
Meas ur em ent of t he M oist ur e Cont ent of P ET F iber (us ing aq u artz cru cibl e)
Specifications
Model TGA-50 TGA-50H TGA-51 TGA-51H
Measurement Temperature Range Ambient to 1000 °C Ambient to 1500 °C Ambient to 1000 °C Ambient to 1500 °C
Weight Measuring Range ±20 mg, ±200 mg ±20 mg, ±200 mg, ±2000 mg
Weight Readability 0.1 μg max.
Sample Mass 1 g including tare 10 g including tare
Atmosphere Control A single flowmeter for the air, inert gas flow, and atmosphere gas is built in.
Dimensions and Weight W173 × D550 × H500 mm, 23 kg W173 × D600 × H540 mm, 25 kg
AC 100 V, 120 V, 220 V, 240 V, AC 100 V, 120 V, 220 V, 240 V, AC 100 V, 120 V, 220 V, 240 V, AC 100 V, 120 V, 220 V, 240 V,
Power Supply
500 VA, 50/60 Hz 1.2 kVA, 50/60 Hz 1 kVA, 50/60 Hz 1.5 kVA, 50/60 Hz
TGA-50 Series
Thermogravimetric Analyzers 3