                                                                 Hyper-MS ™ – a new stan
                                                                 combining core technologies pat





































                 ASDF™ – unparalleled

                       MALDI TOF-TOF
                        ion resolution


              ASDF – Axial Spatial Distribution Focussing is a patented
              Shimadzu technology that enables unparalleled resolution
              in MS/MS mode. Through correction of the axial spatial
              distribution of the ions generated, mass resolution is
              significantly increased and becomes essentially
              independent of the laser power and sample topography.    High- and low-energy
              With ASDF, the MALDI-7090 can achieve mass resolution
              in MS/MS of 10 000 FWHM – unobtainable through                  fragmentation
              pulsed extraction alone.
                                                                   MS/MS fragmentation in the MALDI-7090 is true high-energy
                                                                   CID (HE-CID).
                                                                   The ions enter the CID cell with full 20 keV extraction energy.
                                     y 12
                                                                   Furthermore, collisions with helium rather than heavier gases
               100
               90
                                            PPPPPPPPPPPPPPR        (air or argon) result in the most efficient generation of ions
               80
                                                                   only associated with HE-CID (e.g. side chain fragmentation).
               70
              %Intensity  60 50  w a12  y 12 -2                    MS/MS fragmentation is further enriched by retaining all of
                                                                   the metastable decay ions through the reflectron analyzer,
               40                                                  maximizing the amount of information extracted from each
                                                      w a13
               30
                                                                   sample.
               20
                                                                   All of these features provide the MALDI-7090 with the
               10
                0                                                  highest possible efficiency for metastable and high-energy
                   1190  1200  1210  1220  1230  1240  1250  1260  1270  1280  1290  1300
                                     m/z
                                                                   CID generation and detection.