Page 6 - Shimadzu LC-2030C NT
P. 6

Specifications






                         Model                                              LC-2030C NT
                         P/N                                               S228-65808-58
                         Degassing unit                     Five Lines: Mobile phase 4 + Rinse solution 1 (Volume 400 µL)
                         Pumping method                               Parallel-type double plunger
                         Pulsation                                 ≤ 0.1 MPa (1.0 mL/min,10 MPa, Water)
                         Flow rate setting range                         0.0001 to 10 mL/min
                                                   ≤ ±1% or ±2 µL/min, below whichever is greater (0.01 to 2 mL/min, Specified conditions)
                         Flow rate accuracy
                                                                  ≤ ±2% 2 to 5 mL/min, Specified conditions)
                         Flow rate precision                 ≤ 0.06 %RSD or ≤ 0.02 minSD, below whichever is greater
                       Pump  Configuration                          Four-solvent low-pressure gradient
                         Gradient / range of set concentrations         0 to 100%, in 0.1% steps
                         Gradient / concentration accuracy   ±0.5% (0.1 to 2 mL/min, 1 to 20 MPa, Specified conditions)
                         Gradient / concentration precision     ±0.1% (1 mL/min, 10 MPa, Specified conditions)
                                                                Packed column side 34 MPa (0.0001 to 5 mL/min)
                         Maximum pressure                                     22 MPa (5.0001 to 10 mL/min)
                                                                Slide-in column side 7 MPa (0.0001 to 0.3 mL/min)
                                                                  approximately 615 µL (Packed column side)
                         System Delay Volume
                                                             approximately 620 µL (Slide-in column side) (Option: none)
                         Injection method                             Total-volume sample injection
                         Injection volume accuracy                       ±1% (50 µL, N = 10)
                         Injection volume setting range              0.1 to 100 µL (Option:0.1 to 50 µL)
                                                              RSD ≤ 0.20% (5.0 to 100 µL), RSD ≤ 0.25% (2.0 to 4.9 µL)
                       Autosampler  Cross-contamination        RSD ≤ 0.5% (1.0 to 1.9 µL), RSD ≤ 1.0% (0.5 to 0.9 µL)
                         Injection volume reproducibility
                                                                   0.0025% (Caffeine, Specified conditions)
                         Injection cycle time                        Min. 14 sec (Specified conditions)
                         Samples for processing                 336 (1 mL), 216 (1.5 mL), 112 (4 mL), 4 (MTP/DWP)

                         Sample cooler                                       4 to 45°C
                         Injection linearity                      ≥0.9999 (1 to 100 µL, Specified conditions)
                         Heating and cooling method                   Forced air circulation method
                         Containable column size             Slide-in column    Packed column (1 pieces at 10 to 25 cm)
                       Column Oven  Temperature control range  Setting range 4 to 60°C (To 50°C for slide-in column)
                                                                      Room temperature −5 to 60°C,

                                                                              ±0.1°C
                         Temperature control precision
                         Temperature stability                        ±0.8°C (Specified conditions)
                         Flow rate switching valve                         Not available



















        6