Page 4 - Shimadzu GC-2010 Plus
P. 4

Advanced Flow Technology








                                                                                                                                                                                    Multi-Switching Analysis of Essential Oil
                                                                                                                                                                                    1st Column : MEGA SE-52 0.25 x 25 m df=0.25 μm
                                                                                                                                                                                    Oven temp : 50 °C - 280 °C (3 °C/min)
                                                                                                                                                                                    Injector  : 250 °C
                                                                                                                                                                                    Split ratio  : 1 : 100
                                                                                                                                                                                    Monitor FID : 290 °C
              Advanced Flow Technology is Shimadzu´s solution to provide enhanced separation                                                                                                   H2: 50 mL/min, Air: 400 mL/min, Make-up: 0 mL/min
              power and operational efficiency for applications with complex sample matrices.                                                                                       Switching  : 8 times
              This enhanced capability is based on the high precision Advanced Flow Control                                                                                         2nd Column : MEGA DetTBuSililBeta 0.25 x 25 m df=0.25 μm
                                                                                                                                                                                    Oven temp : 45 °C (12.00 min) - 180 °C (2 °C/min)
              (AFC) of the GC-2010 Plus.


              Advanced Flow Technology for high-performance separation
                                                                                                                                                                                    1st GC Chromatogram
                                                                                                                                                                                    8 compounds are switched onto the 2nd column.
                                                                          Multi-dimensional
                                                                                                                                                                                    2nd GC Chromatogram
                                                                          GC/GCMS System
                                                                                                                                                                                    8 compounds are separated into 15 peaks.
                                                                          MDGC/GCMS-2010
                                                                                                                                                                                                                   Data by
                                                                                                                                                                                                                   Universita degli Studi di Messina
                                                                                                                                                                                                                   Prof. Luigi Mondello
                                                                                                                                                                                                                   Alessandro Casilli
                                                                                                                                                                                                                   Peter Quinto Tranchida
                                                                                                                                                                                                                   Danilo Sciarrone


                                                                                                                                       Chromatogram Obtained with 1 st Column  Reproducibility of 1st Column
                                                                                                                                              Chromatogram
                                                                                                                                                                                          Area        Retention Time
                                                                                                                                                           Heart-cut positions
                                                                                                                                                                              Compound  MeOH  tAmOH  MeOH  tAmOH
                                                                                                                                                                                1     174308  435368  3.008  12.712
                                                                                                                                                                                2     171031  427495  3.009  12.712
                                                                                                                                                                                3     173117  433085  3.009  12.712
         A multi-dimensional GC/GCMS system performs separations using two columns that have                                                                       Without      4     174715  438180  3.009  12.712
                                                                                Switching                                                                          heart-cutting
         different chromatographic selectivity. When components of interest are insufficiently   AuxPress                                                                       5     172315  430744  3.008  12.712
                                                                                flow diagram       Monitor Detector
         separated on the first column, they can be selectively introduced ("heart-cut") to a second                                                                            6     174633  436304  3.008  12.712
         chromatographic column with different selectivity. This enables enhanced chromatographic                                                                   With        7     175269  439073  3.009  12.712
         separation that cannot be attained in conventional single-column analysis. In addition, the   Detector  2 nd  1 st                                         heart-cutting  8  175863  441410  3.008  12.711
         precise flow-switching technology, which is supported by a high-precision digital flow                                                                                 9     172717  430224  3.008  12.712
         controller, ensures heart-cut analysis with the high level of reproducibility demanded of                                                                              10    172002  435496  3.008  12.712
                                                                                           Flow diagram of Standby mode                                                               173597  434738  3.008  12.712
         complex capillary GC separations.                                                                                                                                     Average
                                                                                                AuxPress
                                                                                                                                                                               STDEV  1582   4347   0.00052  0.00032
                                                                                                   Monitor Detector
                                                                                                                                                                               CV(%)  0.912  1.000  0.017  0.002
                                                                                 Detector
                                                                                          2 nd          1 st                           Chromatogram Obtained with 2 nd Column   Reproducibility of 2nd Column
                                                                                                                                                                                                Area                      Retention Time
                                                                                            Flow diagram of Cut mode                                                          Compound  Acetone  IPA  ETBE  Bz  TAME  nBuOH  Acetone  IPA  ETBE  Bz  TAME  nBuOH
                                                                                                                                                                                1   241902  258122  322917  584033  302735  341100  5.515  5.880  11.563  16.501  18.137  18.733
                                                                                                                                                                                2   237073  253283  316835  572472  296948  334684  5.516  5.880  11.565  16.501  18.136  18.734
         • Standby Mode                  • Cut Mode                         Multi-Deans’ Switching                                                                              3   240294  256361  321149  580369  301337  339373  5.516  5.881  11.565  16.503  18.138  18.733
                                                                            In the past, multidimensional GC has been                                                           4   242492  259231  324511  587335  304680  343516  5.515  5.881  11.565  16.502  18.136  18.732
                                                                            accomplished using a switching mechanism known as                                                   5   238727  254946  318735  576768  299044  337067  5.515  5.880  11.563  16.500  18.136  18.733
                                                                            Deans’ Switch. However, this system results in such                                                 6   242091  258606  323204  584763  303246  342028  5.515  5.880  11.563  16.503  18.137  18.732
                                                                            problems as a reduced recovery (sample loss) and                                                    7   243402  260304  325211  588083  305140  344235  5.515  5.880  11.563  16.502  18.135  18.731
                                                                            fluctuations in retention time after column switching.                                              8   241572  260384  326286  588626  301903  337481  5.515  5.880  11.563  16.501  18.135  18.732
                                                                            The MDGC/GCMS-2010 system incorporates multi-                                                       9   239256  255374  319324  576796  299153  336898  5.515  5.880  11.563  16.501  18.135  18.733
                                                                            Deans switching, a new mechanism that significantly                                                 10  238508  254885  319167  576955  299617  337750  5.515  5.880  11.564  16.502  18.137  18.730
                                                                            reduces the likelihood of fluctuations in the retention                                            Average  240532  257150  321734  581620  301380  339413  5.515  5.880  11.564  16.502  18.136  18.732
                                                                            times of components eluted after column switching,                                                 STDEV  2067.56  2506.10  3159.04  5706.12  2662.17  3170.38  0.00037  0.00049  0.00067  0.00076  0.00108  0.00106
                                                                            even when column switching is performed several times.                                              CV   0.860  0.975  0.982  0.981  0.883  0.934  0.007  0.008  0.006  0.005  0.006  0.006




   4                                                                                                                                                                                                                                              5
   1   2   3   4   5   6   7   8   9