                 Specifications                                                                                         AIM-9000
               Hardware
                 Measurement Mode  Transmission/reflection/ATR     Options       Wide-field camera
                 Optical System  15× Cassegrain objective                       (independent illumination from four directions)
                               15× Cassegrain condenser                         ATR objective (Ge)
                 MCT Detector  Wavenumber range: 5,000 to 700 cm -1  (narrow band)  ATR pressure sensor
                                          5,000 to 650 cm -1  (middle band)     Grazing angle objective (GAO)
                               Liquid nitrogen monitoring function              Micro vise holder
                 TGS Detector  Wavenumber range: 4,600 to 400 cm -1             Infrared polarizer
                 (Option)      Automatically switches between detectors, if equipped with   Visible polarizer
                               multiple detectors                 Installation Conditions  Environmental requirements for performance warranty:   Infrared Microscope
                 Supports for   Auto Focus                                      15°C to 30°C with max. 70%RH (and no condensation)
                 Observation and   Automatic adjustment function for Condenser mirror  Site requirements:
                 Measurements  Automatic ATR measurement (with purchase of pressure sensor)  10°C to 35°C with max. 70%RH                AIM-9000
                                                                                (and no condensation)
                                                                                Or max. 60%RH when 30°C is exceeded
                                                                  Power Voltage  100/120/220/230/240 V AC
                                                                  Power Requirements  85 VA

               Software
                 Computer Requirements  OS: Microsoft ®  Windows ®  7 Professional 64bit  Data Processing  • Baseline correction (zero/3-point/multi-point)
                                     Microsoft ®  Windows ®  10 Pro 64bit        • Advanced ATR correction
                               RAM: 8 GB or more                                 • Atmospheric correction
                               Monitor resolution: Min. 1920 × 1080 pixels       • Kramers-Kronig analysis
                 Measurement Support  Automatic aperture setting                 • Differential analysis
                 Functions     (automatic contaminant recognition system)        • Mapping chemicals based on calculation formula
                               Zoom function (max. 330× digital zoom)            • Mapping chemicals based on principal component analysis
                               Multi-image tiling                                • Mapping chemicals based on similarity to reference spectra
                               Up to 60 aperture settings can be specified on microscope   • Spectral search
                               or tiled images.                                  • Pasting spectra and images into other applications
                Validation     Japanese Pharmacopoeia
                               European Pharmacopoeia
                               Chinese Pharmacopoeia
                               Confirmation of aperture size
                Option         Mapping program


                 Examples of Combined Systems
















                                                                                    ™
                                    ™
                            IRTracer -100 +AIM-9000                       IRAffinity -1S+AIM-9000


                        This product is certified as Shimadzu's Eco-Products Plus.
                        Reduced power consumption by 55% compared with conventional Shimadzu's products.




               LabSolutions,IRTracer, and IRAffinity are trademarks of Shimadzu Corporation.
               Microsoft and Windows are registered trademarks of Microsoft Corporation in the United States and/or other countries.






                                                 For Research Use Only. Not for use in diagnostic procedures.
                                                 This publication may contain references to products that are not available in your country. Please contact us to check the availability of
                                                 these products in your country.
                                                 Company names, products/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation,
                                                 its subsidiaries or its affiliates, whether or not they are used with trademark symbol “TM” or “®”.
                                                 Third-party trademarks and trade names may be used in this publication to refer to either the entities or their products/services, whether
                                                 or not they are used with trademark symbol “TM” or “®”.
                                                 Shimadzu disclaims any proprietary interest in trademarks and trade names other than its own.
         www.shimadzu.com/an/                    The contents of this publication are provided to you “as is” without warranty of any kind, and are subject to change without notice.
                                                 Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.
                                                                                                   © Shimadzu Corporation, 2021
                                                                                           First Edition: May 2016   3655-01112-PDFIT
   11   12   13   14   15   16