Page 3 - Kubota - 8730
P. 3

Control panel


                                “TEMP LIMIT” lamp  “TEMP” knob  “SPEED” knob  “TIME” display
                                  “TEMP” display   “SPEED” display  “TIME” key  “TIME” knob











                                      “STEP” key    “MEMORY” and   “PROGRAM” key   “OPEN” key
                                                    “ACCEL/DECEL”
                                         “ACCEL” key  display   “NORMAL” key  “STOP” key
                                           “DECEL” key  “PROGRAM” selection key  “START” key

















          Table of rotors, buckets and tube racks for Model 8730/8622                 Only 1 type of rotor is applicable per unit.
                                                             8730            8622                      Tube Rack
          Rotor Name   Nominal Capacity   Tube size (mm)  Maximum Speed Maximum RCF Maximum Speed Maximum RCF  Bucket  Adaptor
                                           diam. x length                                  Code No. *2
                                                          rpm     xg      rpm     xg                    Code No.
          High capacity swinging bucket rotor
                         4x1,000mL       97〜98x135〜196 (176)  4,000  4,150  3,400  3,000                 —
                     4x700mL stainless steel  97x138     4,000   4,150   3,400   3,000                   —
                     4x500mL conical bottle  *1  96x147  4,000   4,150   3,400   3,000                  431124  *1
           RS-4000   20x50mL conical tube    30x117      4,000   4,130   3,400   2,990  053-5980・053-0010  055-0310
                      16x50mL glass tube  27〜35.8x93〜117  4,000  4,020   3,400   2,910                 055-0330
                     48x15mL conical tube    17x121      4,000   4,110   3,400   2,970                 055-0300
                    4x400mL MAP blood bag      —         4,000   4,080   3,400   2,950                 055-6170L *3
          Multi-purpose swinging bucket rotor
                      130x15mL glass tube  15〜17.2x70〜110  3,000  1,760  3,000   1,760                 055-0050
                      120x15mL glass tube  *4  15〜17.2x70〜110  3,000  1,760  3,000  1,760              055-0060
           RS-1950   36x50mL conical tube    30x117      3,000   1,760   3,000   1,760  equipped with rotor  055-0080
                    6x175/200/225mL conical tube *5  61x118/61x130/61x137  3,000  1,760  3,000  1,760  055-0090
                          6x250mL         60〜62x80〜136   3,000   1,760   3,000   1,760                 055-0090


            Rotor has prescribed operating life cycle, after that the rotor must be replaced.
            For more details, please contact us.
          ✽ The number of times allowed for autoclaving is limited. For more details, please contact us.
          ✽1  Conical bottles made by Corning (No. 431123) require cushions (No. 431124)
          ✽2  RS-4000 requires either bucket Code No. 053-5980 (sealing cap not available) or bucket 053-0010 (sealing cap available).
          ✽3  Inner cup 055-6170L protects blood bags during rotation and facilitates replacement and removal.
          ✽4  Tube rack has two compartments, well-designed for easy handling.
          ✽5  175/225mL conical tubes made by BD Falcon (No. 352076/352075) require cushions (No. 352090).
              200mL conical tubes by Nunc (No. 376813) require cushion adaptors (No. 377585).













          RS-4000 swinging bucket rotor        Bucket                  RS-1950 Swinging bucket rotor
          (with bucket Code No.053-0010)                               (with bucket Code No.055-0060)
                                      Code No.053-5980 Code No.053-0010
                                      (sealing cap is not   (sealing cap is available)
                                      available)
   1   2   3   4